Model 3451
4–5
Figure 4–3. Standard Count Results
If the standard count
fails
and it has been
less than a month
since the last standard count, but the count was performed
correctly and the
failure is less than 5%
, press the
〈
Yes
〉
button.
If the
failure is more than 5%
, press the
〈
No
〉
button. Ensure
that the gauge is properly positioned on the reference standard
block (Figure 4–2). Verify that the standard count site meets the
criteria listed on page 4–2. Take another standard count and
accept it if it fails by less than 5%. If the standard count fails
four times, perform a stat test (see page C–2) and a drift test
(see page C–5), and then contact your Troxler representative.
After the standard count passes, the operator can take measurements
with the gauge. When not taking readings, always keep the source
rod in the
SAFE
position. For added operator safety, the source rod
automatically retracts to the
SAFE
position when the gauge is lifted
using the handle.
4
. US
ING
T
HE
G
AUG
E
Содержание Enhanced RoadReader Plus 3451
Страница 7: ...Model 3451 vii NOTES...
Страница 17: ...xvi NOTES...
Страница 29: ...1 12 NOTES...
Страница 33: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Страница 85: ...4 22 NOTES...
Страница 106: ...Model 3451 6 3 Figure 6 1 Offset Menu 6 OFFSETS CALIBRATION...
Страница 114: ...Model 3451 6 11 Figure 6 10 Moisture Offset Erase Prompt 6 OFFSETS CALIBRATION...
Страница 127: ...6 Figure 6 25 Enter File Name Figure 6 23 Input True Density Figure 6 24 Thin Layer Special Calibration Results...
Страница 131: ...6 NOTES...
Страница 141: ...7 10 Figure 7 5 Project Data Spreadsheet...
Страница 145: ...7 14 Figure 7 8 Project Setup Display Figure 7 9 Standard Counts Display...
Страница 147: ...7 16 Figure 7 10 Project Status Display Figure 7 11 Source Decay Display...
Страница 149: ...7 18 Figure 7 12 Stat Test Display Figure 7 13 Drift Test Display...
Страница 153: ...7 22 NOTES...
Страница 166: ...Model 3451 Appendix A 13 Figure A 3 Cleaning the Sliding and Fixed Blocks...
Страница 174: ...Model 3451 Appendix A 21 NOTES...
Страница 175: ......
Страница 183: ...Appendix B 8 NOTES...
Страница 187: ...Figure C 5 Review Stat Test Appendix C 4 Figure C 3 View Stat Counts Figure C 4 Stat Test Results...
Страница 221: ...Appendix D 6 NOTES...
Страница 223: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 225: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 227: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 229: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 235: ...Appendix G 4 NOTES...
Страница 246: ...Model 3451 Warranty 1 NOTES WARRANTY...