Appendix C–10
PDA CANNOT COMMUNICATE WITH THE GAUGE
✓
Gauge is not turned on.
✓
PDA is not within 4.6 m (15 ft) of the gauge.
✓
Object in the way of the transmitter/receiver.
✓
Check that the PDA is the one shipped with the gauge. Before
leaving the Troxler factory, the transceivers on the gauge and
PDA are programmed with a unique, matching address. The
PDA can only communicate with the matching gauge.
✓
Verify the PDA’s Bluetooth communications configuration as
described on page 3–14.
✓
Verify the PDA’s ActiveSync configuration as described on
page 3–13.
NO DENSITY READINGS
✓
The most likely reason for no density readings is an electronic
problem, such as a failure of the detector preamplifier.
However, there is a remote possibility that the radioactive
source could have become dislodged or lost. Refer to the
section on page A–9 for instructions on
inspecting the source rod.
GAUGE FAILS STANDARD COUNTS
✓
Ensure the source rod opening on the bottom of the gauge is
completely closed or covered by the tungsten sliding block. If
any opening is visible, the sliding block should be cleaned.
Refer to page A–11 for instructions. If the sliding block still
does not close completely, contact the nearest Troxler Service
Center.
✓
Ensure that the standard count site is asphalt, concrete, or
compacted soil at least 4 in. (10 cm) thick.
✓
Check for 10 ft (3 m) clearance around gauge.
✓
Check that any sources are at least 33 ft (10 m) away.
Содержание Enhanced RoadReader Plus 3451
Страница 7: ...Model 3451 vii NOTES...
Страница 17: ...xvi NOTES...
Страница 29: ...1 12 NOTES...
Страница 33: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Страница 85: ...4 22 NOTES...
Страница 106: ...Model 3451 6 3 Figure 6 1 Offset Menu 6 OFFSETS CALIBRATION...
Страница 114: ...Model 3451 6 11 Figure 6 10 Moisture Offset Erase Prompt 6 OFFSETS CALIBRATION...
Страница 127: ...6 Figure 6 25 Enter File Name Figure 6 23 Input True Density Figure 6 24 Thin Layer Special Calibration Results...
Страница 131: ...6 NOTES...
Страница 141: ...7 10 Figure 7 5 Project Data Spreadsheet...
Страница 145: ...7 14 Figure 7 8 Project Setup Display Figure 7 9 Standard Counts Display...
Страница 147: ...7 16 Figure 7 10 Project Status Display Figure 7 11 Source Decay Display...
Страница 149: ...7 18 Figure 7 12 Stat Test Display Figure 7 13 Drift Test Display...
Страница 153: ...7 22 NOTES...
Страница 166: ...Model 3451 Appendix A 13 Figure A 3 Cleaning the Sliding and Fixed Blocks...
Страница 174: ...Model 3451 Appendix A 21 NOTES...
Страница 175: ......
Страница 183: ...Appendix B 8 NOTES...
Страница 187: ...Figure C 5 Review Stat Test Appendix C 4 Figure C 3 View Stat Counts Figure C 4 Stat Test Results...
Страница 221: ...Appendix D 6 NOTES...
Страница 223: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 225: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 227: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 229: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 235: ...Appendix G 4 NOTES...
Страница 246: ...Model 3451 Warranty 1 NOTES WARRANTY...