Model 3451
6–5
MOISTURE OFFSET
The moisture offset (
k
) is determined by comparing the
true
moisture
of a laboratory sample with the gauge moisture reading.
The Model 3451 can derive the moisture offset either by taking a
moisture measurement and comparing the reading to the true
moisture entered by the operator, or by comparing measured and
true values entered by the operator. To access the moisture offset
functions, press the
〈
Moisture
〉
button on the
Offset Menu
(see
Figure 6–1). The PDA displays the
Moisture Offset Select
screen shown in Figure 6–4.
New Moisture Offset from Gauge Reading
To derive the moisture offset by first taking a moisture reading, the
operator must take a reading with the gauge and enter the true
moisture derived by an oven dry or similar method. The gauge can
store one
partial offset
, which allows the operator to take a reading
and enter the true moisture at different times.
NOTE
When using destructive methods such as drilling cores
or sample removal for true moisture measurement, take
gauge readings before removing samples.
Figure 6–4. Moisture Offset Select
6.
O
F
F
S
E
T
S
&
C
A
L
IB
R
AT
IO
N
Содержание Enhanced RoadReader Plus 3451
Страница 7: ...Model 3451 vii NOTES...
Страница 17: ...xvi NOTES...
Страница 29: ...1 12 NOTES...
Страница 33: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Страница 85: ...4 22 NOTES...
Страница 106: ...Model 3451 6 3 Figure 6 1 Offset Menu 6 OFFSETS CALIBRATION...
Страница 114: ...Model 3451 6 11 Figure 6 10 Moisture Offset Erase Prompt 6 OFFSETS CALIBRATION...
Страница 127: ...6 Figure 6 25 Enter File Name Figure 6 23 Input True Density Figure 6 24 Thin Layer Special Calibration Results...
Страница 131: ...6 NOTES...
Страница 141: ...7 10 Figure 7 5 Project Data Spreadsheet...
Страница 145: ...7 14 Figure 7 8 Project Setup Display Figure 7 9 Standard Counts Display...
Страница 147: ...7 16 Figure 7 10 Project Status Display Figure 7 11 Source Decay Display...
Страница 149: ...7 18 Figure 7 12 Stat Test Display Figure 7 13 Drift Test Display...
Страница 153: ...7 22 NOTES...
Страница 166: ...Model 3451 Appendix A 13 Figure A 3 Cleaning the Sliding and Fixed Blocks...
Страница 174: ...Model 3451 Appendix A 21 NOTES...
Страница 175: ......
Страница 183: ...Appendix B 8 NOTES...
Страница 187: ...Figure C 5 Review Stat Test Appendix C 4 Figure C 3 View Stat Counts Figure C 4 Stat Test Results...
Страница 221: ...Appendix D 6 NOTES...
Страница 223: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 225: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 227: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 229: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 235: ...Appendix G 4 NOTES...
Страница 246: ...Model 3451 Warranty 1 NOTES WARRANTY...