6–
SPECIAL CALIBRATION
The
Special Calibration
features allow the operator to recalibrate
Model 3451 soil and thin layer readings. Special calibrations are
specific for the test material and measurement depth. To access
these features, press the
〈
Change Settings
〉
button on the
RoadReader Plus main menu (see Figure 3–11 on page 3–18). The
PDA displays the
Change Settings
menu shown in Figure 5–1 on
page 5–2.
In
Soil
mode, the PDA can activate a special calibration for each
source rod depth.
NOTE
A soil special calibration is active only at the source rod
depth at which it was created. For proper operation,
create a soil special calibration for each required
measurement depth.
Before performing a special calibration, prepare the sample area for
testing as described in
SOIL SPECIAL CALIBRATION
The factory calibration provides accurate results for the majority of
materials encountered in the field. However, there are situations
when varying material compositions and/or texture could affect the
gauge accuracy.
The
Soil Special Calibration
function is used to recalibrate the
Model 3451 gauge for material characteristics not covered by the
factory calibration. The “true density” can be determined by
carefully measuring samples of a known material using an alternate
method in a laboratory or in the field. This function is activated only
in
Soil
mode.
To access the
Soil Special Calibration
function, press the
〈
Special Calibration
〉
button on the
Change Settings
menu
(see Figure 5–1). The PDA displays the
Special Calibration
Menu
shown in Figure 6–13. The following sections provide details
on the
Special Calibration
function.
Содержание Enhanced RoadReader Plus 3451
Страница 7: ...Model 3451 vii NOTES...
Страница 17: ...xvi NOTES...
Страница 29: ...1 12 NOTES...
Страница 33: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Страница 85: ...4 22 NOTES...
Страница 106: ...Model 3451 6 3 Figure 6 1 Offset Menu 6 OFFSETS CALIBRATION...
Страница 114: ...Model 3451 6 11 Figure 6 10 Moisture Offset Erase Prompt 6 OFFSETS CALIBRATION...
Страница 127: ...6 Figure 6 25 Enter File Name Figure 6 23 Input True Density Figure 6 24 Thin Layer Special Calibration Results...
Страница 131: ...6 NOTES...
Страница 141: ...7 10 Figure 7 5 Project Data Spreadsheet...
Страница 145: ...7 14 Figure 7 8 Project Setup Display Figure 7 9 Standard Counts Display...
Страница 147: ...7 16 Figure 7 10 Project Status Display Figure 7 11 Source Decay Display...
Страница 149: ...7 18 Figure 7 12 Stat Test Display Figure 7 13 Drift Test Display...
Страница 153: ...7 22 NOTES...
Страница 166: ...Model 3451 Appendix A 13 Figure A 3 Cleaning the Sliding and Fixed Blocks...
Страница 174: ...Model 3451 Appendix A 21 NOTES...
Страница 175: ......
Страница 183: ...Appendix B 8 NOTES...
Страница 187: ...Figure C 5 Review Stat Test Appendix C 4 Figure C 3 View Stat Counts Figure C 4 Stat Test Results...
Страница 221: ...Appendix D 6 NOTES...
Страница 223: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 225: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 227: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 229: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 235: ...Appendix G 4 NOTES...
Страница 246: ...Model 3451 Warranty 1 NOTES WARRANTY...