Model 4590
4–19
2
2
.
.
Insert the moisture probe into the prepared hole. The
hole must be at least 6 inches deep. The probe baseplate
must have contact with the soil surface. The probe must
fit tightly in the hole and have contact with the soil
material on all sides. You should feel some resistance
when placing it in the hole.
3
3
.
.
If the Moisture Input is set to External Sensor (default
setting). The gauge prompts for this measurement .
4
4
.
.
Once the probe moisture measurement (M) is displayed
on the EGauge screen and has settled on value (3-4
seconds), press Enter to accept the reading and see the
final result for the test.
Entering a Moisture Value from Another
Method
1
1
.
.
If the Moisture Input is set to Manual Input, the gauge
displays:
2
2
.
.
Press
〈
YES
〉
to input the moisture percentage:
3
3
.
.
Enter the moisture percentage using the keypad and press
〈
ENTER/START
〉
.
Prepare Moisture
Measurement?
Press <ENTER> to
take a reading.
Do you know the
Moisture content of
the soil?
Press <YES> or <NO>
%Moisture Input
0.00
Press <ENTER>
Содержание EGauge 4590
Страница 9: ......
Страница 18: ...xvi NOTES...
Страница 24: ...1 6 Figure 1 4590 Gauge and Standard Accessories...
Страница 26: ...1 8...
Страница 70: ......
Страница 94: ...6 4...
Страница 104: ......
Страница 112: ...8 10 Complete Pairing on the Bluetooth enabled device A Confirmation screen displays once pairing is successful...
Страница 164: ......
Страница 168: ......
Страница 171: ...Model 4590 F 3 Standard Count Log Gauge Serial Number Date DS BGD Date DS BGD...
Страница 174: ......
Страница 186: ...Notes NOTES...