Model 4590
4–9
Figure 7. Gauge Position on Soil
1
1
2
2
.
.
If the first attempt fails, check items listed below and retake
standard count. Press
〈
STD
〉
for the display:
NOTE
Manual mode or auto depth sensor detects correct
position. After count completion, the display is:
1
1
3
3
.
.
The
P
indicated to the right of the percentage figures
indicates that the new counts are within the 1% of the
acceptance limits. If the percentages are not within
these
Standard Count
DS= xxxx BGD= xxxx
Take New Count?
<YES> or <NO>
DS= xxxx xx.x% (P/F)
BGD= xxx
Do You Want to
Use the New Std.?