![Troxler EGauge 4590 Скачать руководство пользователя страница 134](http://html1.mh-extra.com/html/troxler/egauge-4590/egauge-4590_manual-of-operation-and-instruction_1172540134.webp)
A-6
Error Messages
The following table displays a list of the gauge error messages,
as well as the possible causes and recommended solutions for
each message.
Table 3. Error Messages
Error Message
Possible Causes
Solution
Bad Checksum!
Default Settings
Restored
Gauge settings and/or
project data stored in
memory have been
corrupted and default
(factory) settings have
been restored.
Set up user preferences
(such as measurement
mode, count time, and so
on) as described in Chapter
3.
Cannot Close File
after Writing.
An error occurred when
attempting to store data
to a USB storage device.
Try to store the data again.
If error recurs, try a
different USB storage
device.
Cannot Initialize
Media
An error occurred when
trying to access the USB
storage device.
Check that the USB storage
device is installed properly.
If error recurs, try a
different USB storage
device.
Cannot Open
Printer
An error occurred when
trying to communicate
with a USB printer.
Check that the printer is
powered on and is properly
connected to the USB port.
Ensure that the printer is
compatible with the gauge.
Cannot Write to
the File
An error occurred when
attempting to store data
to a USB storage device.
Try to store the data again.
If error recurs, try a
different USB storage
device.
Cannot Write to
the Printer
An error occurred when
trying to communicate
with a USB printer.
Check that the printer is
powered on and is properly
connected to the USB port.
Ensure that the printer is
compatible with the gauge.
Содержание EGauge 4590
Страница 9: ......
Страница 18: ...xvi NOTES...
Страница 24: ...1 6 Figure 1 4590 Gauge and Standard Accessories...
Страница 26: ...1 8...
Страница 70: ......
Страница 94: ...6 4...
Страница 104: ......
Страница 112: ...8 10 Complete Pairing on the Bluetooth enabled device A Confirmation screen displays once pairing is successful...
Страница 164: ......
Страница 168: ......
Страница 171: ...Model 4590 F 3 Standard Count Log Gauge Serial Number Date DS BGD Date DS BGD...
Страница 174: ......
Страница 186: ...Notes NOTES...