A-8
Error Message
Possible Causes
Solution
Moisture Offset
Exceeds Limits!
The moisture offset is
outside the accepted
range.
Enter a new moisture
offset.
-nan
Gauge cannot calculate
results.
Take new standard and
background counts. Ensure
the source rod position
matches what is shown on
the screen. If problem
continues, contact the
Troxler Service Center.
No active project! Operator tried to store
measurement data, but
no project is active.
Select an existing project or
create a new project and
select as active.
Proctor Target
Exceeds Limits!
The target value must be
between 80 and 170 pcf.
The operator has
entered a value outside
this range.
Enter a target value
between 80 and 170
pcf.
Project has No
Data!
Operator attempted to
view a project with no
data.
Select the correct project
when attempting to view.
Warning!!
Time to Reach
Precision too
high
When using the
Precision
function, the
time required to reach
the selected precision
value exceeds 60
minutes.
Enter a different precision
value or disable the
Precision function.
Error 13
System is unable to
initialize.
Issue with Gamma Detector.
Contact Troxler Service
Center
Error 64
System is unable to
initialize.
Issue with Gamma Detector.
Contact Troxler Service
Center
Remote Keypad
Failure
If remote START button is
active for more than 20
sec “Remote START
button failure! Displays.
Press YES to ignore and
continue. The same
changes apply to remote
ESC button.
Содержание EGauge 4590
Страница 9: ......
Страница 18: ...xvi NOTES...
Страница 24: ...1 6 Figure 1 4590 Gauge and Standard Accessories...
Страница 26: ...1 8...
Страница 70: ......
Страница 94: ...6 4...
Страница 104: ......
Страница 112: ...8 10 Complete Pairing on the Bluetooth enabled device A Confirmation screen displays once pairing is successful...
Страница 164: ......
Страница 168: ......
Страница 171: ...Model 4590 F 3 Standard Count Log Gauge Serial Number Date DS BGD Date DS BGD...
Страница 174: ......
Страница 186: ...Notes NOTES...