Model 4590
4–7
5
5
.
.
Press
〈
START
〉
to begin taking the 2-minute standard
count.
NOTE
The standard and background standard must each be
within 1% of the average of the last 4 counts
performed and stored. If the standard count average
is greater than 1%, the standard count fails, and an
error code displays as shown below:
6
6
.
.
When count is finished and passed, the first step results
displays:
7
7
.
.
Press
〈
ENTER
〉
to continue:
8
8
.
.
Lower the handle to the background position (first notch
below SAFE). The gauge automatically checks whether the
Step 1 Ct.= xxxx
<ENTER> to continue
Taking
Standard Ct. Step 1
xxx s
Remaining
Error Code#:11
Please Refer
to Users Manual
Press <ENTER>
With Gauge on
Test Site & Source
Rod in Background Pos.
Press <START>