4-6
TRENCH OFFSETS
If the Model 3430 gauge is to be used for moisture or density
measurements in a trench or within two feet (0.6 m) of a large
vertical structure, a trench offset may be required. If used, the trench
offset adjusts all moisture measurements but only the density
measurements from backscatter through four inches (10 cm).
To perform a trench offset:
✓
Take the daily standard count (outside the trench) and record the
density count (
DS
) and moisture count (
MS
)values.
✓
Place the gauge on the reference standard block in the trench the
same distance from the wall as the anticipated readings.
Do not
take another standard count.
✓
Set the count time to four minutes.
✓
With the source rod in the
SAFE
(standard count) position, take
a four-minute count. To start the count, press the
〈
START/ENTER
〉
key.
✓
Record the trench density count (
DC
Trench
) and moisture count
(
MC
Trench
.).
✓
Subtract the daily standard count values from the trench count
values:
Dens Const = (DC
Trench
) – DS
Moist Const = (MC
Trench
) – MS
To enable a trench offset, choose
Trench
from the
Offset
options
under the
Special
functions. The gauge requests the
Dens Const
and
Moist Const
values determined above. The procedure for entering
the values is the same as for moisture and density offsets,
ignoring
the
±
sign on the display.
Содержание 3430-M
Страница 9: ......
Страница 14: ...xii NOTES...
Страница 19: ...Model 3430 1 5 Figure 1 1 Model 3430 Gauge and Accessories INTRODUCTION...
Страница 22: ...1 8 NOTES...
Страница 25: ...Model 3430 2 3 Figure 2 1 Direct Transmission Geometry Figure 2 2 Backscatter Geometry OPERATION THEORY...
Страница 26: ...2 4 Figure 2 3 Backscatter Surface Density Effects Top Layer Effect Curves...
Страница 28: ...2 6 Figure 2 4 Effect of Moisture on Depth of Measurement...
Страница 48: ...3 20 NOTES...
Страница 62: ...4 14 NOTES...
Страница 66: ...5 20 D xxxx M xxxx Use keys Moist R xxxx to view data Moist Avg xxxx to view data...
Страница 88: ...A 14 NOTES...
Страница 111: ...Model 3430 C 17 Figure C 2 3430 Gauge Assembly MAINTENANCE...
Страница 113: ...Model 3430 C 19 Figure C 3 3430 Gauge Base Mechanical Assembly Figure C 4 3430 Source Rod Handle Assembly MAINTENANCE...
Страница 115: ...Model 3430 C 21 Figure C 5 3430 Gauge Preamplifier Assembly MAINTENANCE...
Страница 117: ...Model 3430 C 23 Figure C 6 3430 Gauge Scaler Assembly MAINTENANCE...
Страница 128: ...E 2 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Страница 129: ...Model 3430 E 3 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number STANDARD COUNT LOG...
Страница 130: ...E 4 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Страница 140: ...Index 8 NOTES...
Страница 141: ...Model 3430 Warranty 1 NOTES WARRANTY...