GARBAGE OR
XXXXXX
IS DISPLAYED
✓
Check the standard counts in memory. If the standard counts are
suspect, perform new standard counts. If counts equal zero for
both systems, replace high voltage board (contact the nearest
Troxler Service Center).
✓
Check gauge for water damage. If the gauge is wet, dry the
gauge interior with hairdryer (on low heat) for 3 hours.
✓
Check the calibration constants. They should match the
constants on your calibration data sheet
if your calibration
(Refer to the note on page 5-10 to
✓
If necessary, perform a statistical stability (stat) test, record the
results and contact the nearest Troxler Service Center.
GAUGE TURNS OFF AFTER IT IS TURNED ON
✓
The gauge automatically turns off after five hours if no keys are
pressed. Try to turn the gauge on again.
✓
The gauge may be wet. Do not turn the gauge on until moisture
is removed from gauge interior! Component damage may result.
✓
If the battery is below 3.0 volts, recharge or replace the
batteries.
✓
The scaler may be defective. To test, replace the suspect scaler
with a good scaler.
SHORT BATTERY LIFE AFTER RECHARGING
✓
NiCad batteries may be charged up to 100 full charge-discharge
cycles. The batteries may be reaching end of life cycle - replace.
Note that all information stored in the gauge except the
calibration constants and the chosen language is lost when the
batteries are disconnected.
✓
Charger/adapter may not be supplying full charge – check the
AC outlet and the DC output (12 VDC).
C-4
Содержание 3430-M
Страница 9: ......
Страница 14: ...xii NOTES...
Страница 19: ...Model 3430 1 5 Figure 1 1 Model 3430 Gauge and Accessories INTRODUCTION...
Страница 22: ...1 8 NOTES...
Страница 25: ...Model 3430 2 3 Figure 2 1 Direct Transmission Geometry Figure 2 2 Backscatter Geometry OPERATION THEORY...
Страница 26: ...2 4 Figure 2 3 Backscatter Surface Density Effects Top Layer Effect Curves...
Страница 28: ...2 6 Figure 2 4 Effect of Moisture on Depth of Measurement...
Страница 48: ...3 20 NOTES...
Страница 62: ...4 14 NOTES...
Страница 66: ...5 20 D xxxx M xxxx Use keys Moist R xxxx to view data Moist Avg xxxx to view data...
Страница 88: ...A 14 NOTES...
Страница 111: ...Model 3430 C 17 Figure C 2 3430 Gauge Assembly MAINTENANCE...
Страница 113: ...Model 3430 C 19 Figure C 3 3430 Gauge Base Mechanical Assembly Figure C 4 3430 Source Rod Handle Assembly MAINTENANCE...
Страница 115: ...Model 3430 C 21 Figure C 5 3430 Gauge Preamplifier Assembly MAINTENANCE...
Страница 117: ...Model 3430 C 23 Figure C 6 3430 Gauge Scaler Assembly MAINTENANCE...
Страница 128: ...E 2 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Страница 129: ...Model 3430 E 3 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number STANDARD COUNT LOG...
Страница 130: ...E 4 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Страница 140: ...Index 8 NOTES...
Страница 141: ...Model 3430 Warranty 1 NOTES WARRANTY...