Model 3430
C-15
LEAK TESTING
Unless specified otherwise by your license
, the gauge must be leak
tested at intervals not exceeding 12 months to ensure the integrity of
the radioactive source encapsulation. Sample analysis must be
performed by a licensed laboratory only.
Using the Troxler Model 3880 Leak Test Kit and accompanying
instructions, perform the following procedure:
Ensure that the source rod is in the SAFE position (see
Figure 3-2).
✓
Write the date, gauge model number, and serial number on the
wipe disk.
✓
Remove the control panel from the gauge topshell. Locate the
yellow radiation label on the top surface of the base.
✓
Place two drops of leak test solution on the wipe disk.
✓
Holding the wipe disk with the tongs, wipe the radiation label.
✓
Turn the gauge on its side and locate the opening where the
source rod extends through the gauge base.
✓
Holding the wipe disk with tongs, wipe the area around and
inside the opening where the source rod extends from the gauge
base.
✓
Pack the disk, as instructed, in the envelope and mail to Troxler
Electronic Laboratories, Inc. for analysis.
✓
Secure the gauge properly.
M
AI
NT
E
NANC
E
Содержание 3430-M
Страница 9: ......
Страница 14: ...xii NOTES...
Страница 19: ...Model 3430 1 5 Figure 1 1 Model 3430 Gauge and Accessories INTRODUCTION...
Страница 22: ...1 8 NOTES...
Страница 25: ...Model 3430 2 3 Figure 2 1 Direct Transmission Geometry Figure 2 2 Backscatter Geometry OPERATION THEORY...
Страница 26: ...2 4 Figure 2 3 Backscatter Surface Density Effects Top Layer Effect Curves...
Страница 28: ...2 6 Figure 2 4 Effect of Moisture on Depth of Measurement...
Страница 48: ...3 20 NOTES...
Страница 62: ...4 14 NOTES...
Страница 66: ...5 20 D xxxx M xxxx Use keys Moist R xxxx to view data Moist Avg xxxx to view data...
Страница 88: ...A 14 NOTES...
Страница 111: ...Model 3430 C 17 Figure C 2 3430 Gauge Assembly MAINTENANCE...
Страница 113: ...Model 3430 C 19 Figure C 3 3430 Gauge Base Mechanical Assembly Figure C 4 3430 Source Rod Handle Assembly MAINTENANCE...
Страница 115: ...Model 3430 C 21 Figure C 5 3430 Gauge Preamplifier Assembly MAINTENANCE...
Страница 117: ...Model 3430 C 23 Figure C 6 3430 Gauge Scaler Assembly MAINTENANCE...
Страница 128: ...E 2 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Страница 129: ...Model 3430 E 3 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number STANDARD COUNT LOG...
Страница 130: ...E 4 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Страница 140: ...Index 8 NOTES...
Страница 141: ...Model 3430 Warranty 1 NOTES WARRANTY...