Model 2701-B
3–9
3. ADJUSTI
NG
M
EASUREMEN
TS
The gauge can store the mix calibration in one of four
memory cells
.
Use the numeric keys to store the mix calibration. Note that this will
replace any mix calibration that was previously stored in that cell.
The gauge enables the mix calibration, displays a brief confirmation
message, and returns to the
Setup
menu.
Slope/Intercept
The
Slope/Intercept
function is used to manually enter the slope
and/or intercept for a particular material. These values can be
determined using the
function described on page 3–7, or
can be calculated manually.
To access the
Slope/Intercept
function, press
〈
2
〉
on the
Entry
Method
menu shown on page 3–7. The gauge prompts for the
slope and intercept values. Press
〈
ENTER/START
〉
after each value
is entered. After both values have been entered, the gauge displays:
Slope = #.##
Intercept = ###.#
Do you want to
accept this mix?
To return to the
Setup
menu without saving the mix calibration,
press the
〈
NO
〉
key. To save the mix calibration, press the
〈
YES
〉
key. The gauge requests a mix calibration ID as shown:
Mix Calib. ID
<ALPHA> for Letters
Press <ENTER>
Press the
〈
ALPHA LOCK
〉
key to enable the alphabetic keys on the
gauge. When the alphabetic keys are enabled, the symbol
A
appears
in the upper right of the display, as shown above.