Model 2701-B
3–7
3. ADJUSTI
NG
M
EASUREMEN
TS
NEW MIX CALIBRATION
To determine a new mix calibration or to manually enter the slope
and intercept values, press
〈
5
〉
on the menu shown on page 3–6. The
gauge displays:
-Entry Method-
1. Data Pairs
2. Slope/Intercept
To determine a new mix calibration based on gauge measurements
and values determined by an alternate method, press
〈
1
〉
to access
the
Data Pairs
function. To enter slope and intercept values
manually, press
〈
2
〉
to access the
Slope/Intercept
function. These
functions are described in the following sections.
Data Pairs
To determine a new mix calibration based on gauge measurements
and values determined by an alternate method, press
〈
1
〉
at the
Entry Method
menu shown above. The gauge displays:
Input Reading #: 1
0.0 pcf
Press <ENTER>
Using the numeric and decimal keys, enter the
measured density
(the density as measured by the gauge) at the first test location.
Press the
〈
ENTER/START
〉
key. The gauge displays:
Input Core Value 1:
0.0 pcf
Press <ENTER>
Enter the
actual density
(the density as determined from an alternate
method) and press the
〈
ENTER/START
〉
key.