Parameter identifiers always refer to a certain firmware version. Therefore an
individual parameter reference table exists for every firmware version.
You find the description of the binary commands for the processing of the user
parameters in the “PS Laser Scanner Programmer’s Manual”.
There are different types of parameter:
•
Usual parameters can be set by the user for sensor configuration. Those
parameters are stored in the non-volatile memory of the sensor.
•
Constant parameters
Constant parameters
Constant parameters
Constant parameters (e.g. the serial number) cannot be changed by the
user.
•
Temporary pa
Temporary pa
Temporary pa
Temporary parameters
rameters
rameters
rameters (e.g. the system temperature) will be overwritten if
a device status has been changed. Note that temporary parameters are
typically updated by a self-test during the startup phase. It is
recommended to set temporary parameters only after the self-test has
been finished.
7.5.3
Table of User Parameters
The following table contains all parameters to be set by the user:
ID
ID
ID
ID Default
Default
Default
Default
Name
Name
Name
Name
Description
Description
Description
Description
1
Const: User settings
A version number of the parameter set
2
0
User parameter
password, 0=disabled
The terminal mode can be protected by a password
number.
Enter "0" to disable the password
3
0
Scan mode:
off=0,
normal=1, fast=2,
fine=3
This parameter defines the operating scan mode:
fine/normal/fast.
4
0
Auto-start without
SCAN/GSCN command,
enabled=1
If this parameter is 1, the scanner provides a
continuous output stream of profiles
5
0
Start mode,
0=Standard,
1=Terminal Mode
If this parameter was set to 1, the scanner enters the
Terminal mode after startup.
6
1
Startup message on
serial interface,
enabled=1
If this parameter was set to 1, the startup messages like
the firmware version are not displayed on the serial
interface.
7
1
Red Laser Marker at
startup:
off=0, on=1, auto=2
If this parameter was set to 1, the red laser marker is
switched after startup.
Set to 0 if you wish the red laser marker to be off after
startup.
8
1
Temp: Red Laser
Marker status:
Temporary parameter indicating the current status of