OSCAR G2
Manual
D01-44en201810
Software V.1.1.6
(C)TriOS Mess- und Datentechnik GmbH
Bürgermeister-Brötje-Str. 25
D- 26180 Rastede/Germany
7 / 62
1.2 Warning notices
This sensor is designed for use in science. It should only be used for measuring
aqueous solutions, e.g. river water or sea water.
Sensors made of stainless steel are not made for use in sea water
(corrosion). Only sensors of titanium can be used in sea water.
Sensors which are made of stainless steel have to be cleaned immediately after
exposure to salt water or other corrosion-causing substances (e.g. acids, alkalis,
chlorine-based compounds).
Even Titanium will be affected for example by bromine, hydrofluoric acid and hot
acids. Do not use sensors of titanium in these media.
The sensor
OSCAR
has seals of NBR (nitrile butadiene rubber). On custom request
sealing rings of other materials can possibly be used. Before operation take care that
the assay medium does not damage the seals.
Before operating check the power supply.
OSCAR
has to be operated with voltages
between 12 and 24 VDC.
Don’t use alcohol, benzene, dilutor or other flammable substances to clean or
maintain the equipment. The use of these substances may lead to fire.
Exception:
cleaning of the optical windows with a few drops of acetone and optical
paper from
TriOS Mess- und Datentechnik GmbH
or a lint-free tissue and cleaning
of the cavity with ethanol.
Do not cut, damage nor modify any of the cables. Make sure that no heavy objects
are located on the cable or that the cables are kinked.
Make sure that the cable does not run near hot surfaces.
If the power cable is damaged, it must be replaced only by
TriOS Mess- und
Datentechnik GmbH
or an authorized
TriOS
workshop with an original replacement.