After the measurements the data are downloaded and the calibrated waterbase and
Nigrosin spectra are imported into the database of
msda_xe
.
The calculation of the reflectivity is performed with the averaged spectrum of pure
water measured before the Nigrosin measurement.
The reference Nigrosin measurement is imported from clipboard as described in
chapter 4.2.1.
The reflectivity is calculated.
6.3 Reference spectrum of Nigrosin
The following figure shows an example of the nigrosin reference measurement with an
absorption spectrometer Perkin Elmer Lambda 850.
Fig. 35. Reference spectrum of Nigrosin recorded with Perkin Elmer Lambda 850
Lambda 850: Absorption spectrum of Nigrosin
c = 0.4 mg/l_100 cuvette