The following setting options are available under "Peripherals":
Transceiver:
Specifies the standard electrical connection.
Possible options are:
o
EIA-232 (also RS-232) and
o
EIA-485 (also RS-485)
Protocol:
Specifies the protocol to be used for data.
o
Modbus RTU
o
TriOS data protocol
o
IEEE 488.2 (SCPI)
o
ASCII Output
Baud rate.
Specifies the selectable baud rate, i.e. transmission speed.
Note:
Reduce
the baud rate should there be communication difficulties.
Flow control.
Enables flow control on the software level (XON / XOFF).
Note:
This is
only supported with the
TriOS data protocol
and must be disabled when using
Modbus RTU.
Parity.
Activates the parity check during data transfer. Possible options are:
o
None (disabled)
o
Even
o
Odd
Stop bits.
Specifies the number of stop bits.
Note:
For various Modbus devices, it
may be necessary to set it to "Two" if parity check is to be avoided.
4.1.5 System
The "System" page (Fig. 21) is used to configure general
OSCAR
sensor parameters such
as date and time. Furthermore, it allows downloading the current calibration as a backup.