OSCAR G2
Manual
D01-44en201810
Software V.1.1.6
(C)TriOS Mess- und Datentechnik GmbH
Bürgermeister-Brötje-Str. 25
D- 26180 Rastede/Germany
46 / 62
Fig. 24. Web-Interface: Calibration sub-item Waterbase, settings
We recommend to set the averaging from 5 to 10 and the pause to 60 seconds.
Set the temperature for the ultra pure water and then press the
“calibrate” button (Fig. 24)
.
On the sub-item
“Nigrosin” (Fig. 25) a new measurement of Nigrosin can be taken. You also
can perform an average measurement and you have to set the temperature of the solution.