necessary. Please contact support of
TriOS Mess- und Datentechnik GmbH
4.3 Operation in practice
In the following the practical operation with
OSCAR
is described:
Connect the
OSCAR
via
G2 Interface Box
with your PC or laptop (see section 3.1.3),
open the sensor menu with a web browser and configure the settings (see 4.1.3).
Enter a meaningful comment according to your medium and set the measuring
interval to a value you need.
Fill the cavity with the medium; note the temperature and the salinity of the medium.
If the temperature and salinity of the medium are constant, you can set the values
before starting automatic measurements under the page “measurement” “spectrum”
(see Fig. 14). Confirm them by executing a single measurement.
Enable automatic measurements (see Fig. 18).
Fig. 32. Setting for automatic measurement
After finishing the automatic measurements, open the sensor menu with a web
browser and download the measured data.
If it is necessary to correct the temperature and salinity values of the recorded
spectra, open the dat.file with calibrated spectra. Set the temperature of each
spectrum to the measured temperature of the medium. Also set the value for salinity.
Save the file and import it into the database of the
MSDA_XE
software.
Start the calculation of absorption under consideration of the reflectivity (see chapter