For measurements with
OSCAR
the interval should set ≥ 60s, otherwise the temperature
correction of the LED intensities does not work properly.
The integration time (= IT) should always be in automatic mode or readings
can become unreliable.
The maximum values are 12 h for the interval, 24 for the averaging and 60 s for pause.
If the entered values do not make sense or the spectra cannot be calculated within the given
time, an error message appears like shown in Fig. 19 when activating the automatic mode.
If you want to use the averaging function, set the pause to 60s to avoid
strong warming effects of the
LED’s.