Trident RiX-70 DC Скачать руководство пользователя страница 9

 
 

 

User Manual 

RiX 70 DC

 

 

 

60 kV 

Tooth/Patient 

size 

Digital sensor 

Film type D 

Film type E 

Film type F 

Phosphor sensor 

Adult 

Child  Adult  Child  Adult  Child  Adult  Child 

Adult 

Child 

M

a

xi

lla

 

Molars 

0.200 

0.125  0.630  0.400  0.250  0.160  0.200 

0.125 

0.630 

0.400 

Premolars 

0.160 

0.100  0.500  0.320  0.200  0.125  0.160 

0.100 

0.500 

0.320 

Incisors 

0.125 

0.063  0.400  0.250  0.160  0.100  0.125 

0.080 

0.400 

0.250 

M

a

nd

ib

le

 

Incisors 

0.110 

0.040  0.250  0.160  0.100  0.063  0.080 

0.050 

0.250 

0.160 

Premolars 

0.125 

0.071  0.320  0.200  0.125  0.080  0.100 

0.063 

0.320 

0.200 

Molars 

0.160 

0.100  0.500  0.360  0.200  0.125  0.160 

0.100 

0.500 

0.360 

 

Bite-wing 

0.200 

0.125  0.800  0.500  0.320  0.200  0.250 

0.160 

0.800 

0.500 

 

3.6  Manual selection of Exposure time  

The user can manually modify the pre-set exposure times for any type of receptor,  pressing one of 
the buttons “+” or “-“ from the main keyboard or on the optional keyboard. Times can move of 47 
steps from 0.010 to 2s as per the following table:  
 

0.010 

0.011 

0.012 

0.014 

0.016 

0.018 

0.020 

0.022 

0.025 

0.028 

0.032 

0.036 

0.040 

0.045 

0.050 

0.056 

0.063 

0.0710 

0.080 

0.090 

0.100 

0.110 

0.125 

0.140 

0.160 

0.180 

0.200 

0.220 

0.250 

0.280 

0.320 

0.360 

0.400 

0.450 

0.500 

0.560 

0.630 

0.710 

0.800 

0.900 

1.000 

1.100 

1.250 

1.400 

1.600 

1.800 

2.000 

 

 

 

 

 
 
4  CARE AND MAINTENANCE 
 
4.1 Cleaning  

Always disconnect the line voltage supply before cleaning the unit. Use a mild soap to remove fin-
ger or other dirty marks paying attention not to have liquids enter into the equipment.  
 

 

Clean painted surfaces , accessories and connection cables only by hand with 
a damp cloth and mild detergent, making sure to wipe with a dry cloth ; Do not 
use solvents ( alcohol, benzene , trichlorethylene ), corrosive and abrasive. 
 

 

4.2 Disinfecting  

Parts in touch with the patient must be cleaned with a detergent (such as 2% solution of ammonia) 
making sure not to use solvents or corrosive  disinfectants, which can cause cracks  on the plastic 
covers.  
 

4.3 Maintenance  

Maintenance of RiX- 70 DC to be done regularly by a service technician at least once every 24 
months, in addition to regular checks performed by the operator every year.  
 

 

RiX-70 DC system not contain parts that can be repaired by the user. 
The system must be opened and maintained by qualified personnel. 
 

 

Содержание RiX-70 DC

Страница 1: ...User Manual RiX 70 DC i User Manual RiX 70 DC X Ray Intraoral System Date 19 01 2017 Language English Rev 2 Code UM04 EN 0051 ...

Страница 2: ... 4 1 Cleaning 8 4 2 Disinfecting 8 4 3 Maintenance 8 5 DISPOSING OF OBSOLETE EQUIPMENT 9 6 ELECTROMAGNETIC COMPATIBILITY 9 6 1 Electromagnetic Emissions 9 6 2 Electromagnetic Immunity 10 6 3 Recommended Separation Distances Between Portable and Mobile RF Communication Equipment and RiX 70DC 11 Appendix A Systems 12 Appendix B Icons 14 Appendix C Heating and Cooling Curves 15 Appendix D Alarms 16 A...

Страница 3: ... performances which could cause death injuries or health hazard to Patient and or Operator 1 2 Safety Recommendations Carefully read all the safety instructions to avoid potential hazards that could result in personal injuries or could damage your equipment This table contains important information regarding the system the operator and the patient safety Please read this table very carefully RiX 7...

Страница 4: ... operation of the system by involving a specialist for any maintenance Do not use the device in the presence of vapours or flammable anaesthetic mixtures with air or oxygen or nitrous oxide When using digital sensors if the PC is positioned within the patient environment it must meet the requirements defined by the IEC 60601 1 medical device if located outside it must be in compliance with IEC 609...

Страница 5: ...cm Beam diameter X Ray at 200 mm focus 60 mm Bean limiting device additional Rectangular 36 x45 mm Dose intensity accuracy 50 2 3 X Ray Tube Manufacturer Toshiba Japan CEI Italy Model D 045 OX70 4 Inherent filtration 1 mm Al eq 0 5 mm Al eq Target angle 12 5 16 Filament current Max 3 A 1 5 2 1 A Max filament current 3 V 1 7 3 1 V Anodic thermal capacity 4 3 kJ 7 kJ Max anodic dissipation 100 W 110...

Страница 6: ...u it s necessary to push again Menu button X RAY button Activation pressing this button will be activated on the display a count down during which you can make expositions by pressing the X Ray button Tooth Type selection choose the tooth to be examined lower incisors lower canines lower bicuspids lower molars upper molars bicuspids upper canines upper teeth bitewing System ready green light ON kV...

Страница 7: ... symbol of your selection in the case of analog receptor it is also indicated the sensitivity of the selected film 3 Select the patient size and type of tooth Tooth type touch the symbol corresponding to the tooth you want to x ray the display will show the symbol corresponding to the selection Once you select these parameters RiX 70 DC automatically selects the output time for the type of X Ray t...

Страница 8: ...ds on many aspects as per the user preference contrast brightness etc and per patient mouth anatomy 70 kV Tooth Patient size Digital sensor Film type D Film type E Film type F Phosphor system Adult Child Adult Child Adult Child Adult Child Adult Child Maxilla Molars 0 100 0 063 0 320 0 200 0 125 0 080 0 100 0 063 0 320 0 200 Premolars 0 080 0 050 0 250 0 160 0 100 0 063 0 080 0 050 0 250 0 160 Inc...

Страница 9: ...5 0 028 0 032 0 036 0 040 0 045 0 050 0 056 0 063 0 0710 0 080 0 090 0 100 0 110 0 125 0 140 0 160 0 180 0 200 0 220 0 250 0 280 0 320 0 360 0 400 0 450 0 500 0 560 0 630 0 710 0 800 0 900 1 000 1 100 1 250 1 400 1 600 1 800 2 000 4 CARE AND MAINTENANCE 4 1 Cleaning Always disconnect the line voltage supply before cleaning the unit Use a mild soap to remove fin ger or other dirty marks paying atte...

Страница 10: ...ribed below 6 1 Electromagnetic Emissions The device may cause radio interference and may disrupt the operation of nearby equipment It may be necessary to take mitigation measures such as re orienting or relocating the equipment or shielding the location The system comes with all the cables needed for its operation Using other cables not supplied by the manufacturer or in addition to those provide...

Страница 11: ...al or hospital environment If the user of the device requires continued operation during power mains interruptions the device needs to be powered from an uninterruptible power source Power frequency 50 60Hz magnetic field IEC 61000 4 8 3 A m Test level IEC 60601 1 2 Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environm...

Страница 12: ...distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be determined using the equation applicable to the frequency of the tra...

Страница 13: ...0 DC 12 Appendix A Systems 1 2 7 6 3 4 5 8 Walll mounted version 1 Wall plate 2 Support arm 40 60 80 cms 3 Scissor arm 4 Monoblock Support tie 5 Monoblock 6 Cone collimator 7 Control panel timer 8 Handswitch with cable ...

Страница 14: ...User Manual RiX 70 DC 13 Stand mobile version 1 Mobile stand base 2 Mobile stand column 3 Control unit timer hand switch with cable 4 Scissor arm 5 Monoblock with cone collimator 1 2 3 4 5 ...

Страница 15: ...ended function The product at the end of its useful life can not be thrown in the regular trash with other wastes and is subject to a separate collection Alternating Current N Conductor connected to the neutral point L Connection point to the Power Line Grounding Functional earth connection ON OFF POWER ON Power OFF Electric shock risk Product Reference Serial Number Manufacturing date Manufacture...

Страница 16: ...User Manual RiX 70 DC 15 Appendix C Heating and cooling curves CEI TUBE TOSHIBA TUBE Monoblock cooling curve ...

Страница 17: ...tion does not match the specifications Repeat the configuration of the system Call the tech assistance to replace the CPU board if the problem occurs frequently Error 1 X Ray button pressed No operating system Release the keys eventually pressed turn the system off and on again Consult the Maintenance Manual Error 5 More than one button pressed during the switching on System inoperative Release th...

Страница 18: ...rror 8 Open door The room s door if equipped with the proper sensor has been opened during the phase Ready for X RAY Close the door and restart the operations pressing Menu button Blocking messages These messages identify warnings situations for the patient the operator and other persons located close to the machine For these reason these errors block the machine Display info Description Cause Sol...

Страница 19: ... N Label 3 Scissor Double pantograph Arm Model it s a fixed code that identifies the double pantograph arm E00 03 Serial Number It s an alphanumeric code of 2 letters 6 numbers for instance EA000001 EA identifies the double pantograph arm 00001 identifies the sequence productive number Label 4 Collimator Bean Limiting Device REF E00 04 00 021 it s a fixed code that identifies the rectangular colli...

Страница 20: ...trident dental com info trident dental com IMPORTANT Please do not try to open your unit Only trained and qualified technicians authorized by Trident can remove the covers giving access to the circuits Please contact an authorized Trident distributor responsible for the technical assistance of your device ...

Отзывы: