Trident RiX-70 DC Скачать руководство пользователя страница 17

 
 

 

User Manual 

RiX 70 DC

 

16 

 

 

Appendix D 

Alarms  

 

Display info 

Description 

Cause / Solution 

------ 

The display is completely black 

No  power.  Reset  and  switch  on  again  ;  if  the  error 
persists, please consult the Maintenance Manual 
If the power switch and the indicator lights are switched 

on, probably the main fuse is broken .  
Consult the Maintenance Manual. 

------- 

The display is completely blank 
even  if  the  switch  is  on  (green 
light on) 

Likely failure of the main fuse. Consult the Maintenance 
Manual. 
If the error persists, fatal error of the program memory.  
Call the tech assistance to Replace the CPU board.  

Error 9 

Memory Error EEPROM 1 

Control error of the first memory bank data ; the 
system is operating as it has restored the data from 
the second memory bank . Press " Menu " to continue 
checking. 
Call the tech assistance to replace the CPU board if 

the problem occurs  frequently 

Error 9 

Memory Error EEPROM 2 

Control error of both the first and the second memory 
bank; the system is operating as it has restored the 
data from the program memory. Press " Menu " to 
continue checking . 
Call the tech assistance to replace the CPU board if 

the problem occurs  frequently

 

Error 10 

Non congruent configuration 

The system detected that the system configuration 
does not match the specifications. 
Repeat the configuration of the system. 
Call the tech assistance to replace the CPU board if 

the problem occurs  frequently 

Error 1 

X-Ray button pressed  

No operating system 
Release the keys eventually pressed; turn the system 
off and on again. 
Consult the Maintenance Manual. 

Error 5 

More  than  one  button  pressed 
during the switching on. 

System inoperative. 
Release the buttons; turn the system off. On again. 
If the problem persists, please consult the 
Maintenance Manual. 

 

        

Error during pre-exposure phase  

 

Display info 

Description 

Cause / Solution 

 

Mains voltage too low. 

The  system  supply  voltage  has  dropped  below  the 
minimum level allowed for the security of the system. 

 

Press " Menu " to delete the message . If it reappears or 

if  the  situation  frequently  happens,  verify  that  the 
office’s supply voltage is within the tolerances. 

Error 1 

 
Button 

pressed 

without 

enabling  
 

The system is protected by a hardware protection key 
that  was not correctly positioned before pressing the 
X-Ray button .

 

Release the button,  press the "Menu" button to resume 
the system operation of and turn the security key to the 
enabled position before repeating the exposure. 
Release the button and press the "Menu" button to 
resume operation of the system. 
Enter the state of "Ready for X-Rays" before performing 
the exposure. 
In the remote system.

 

X-Ray button was not simultaneously pressed with   "+" 
button to enable the emission. 

Содержание RiX-70 DC

Страница 1: ...User Manual RiX 70 DC i User Manual RiX 70 DC X Ray Intraoral System Date 19 01 2017 Language English Rev 2 Code UM04 EN 0051 ...

Страница 2: ... 4 1 Cleaning 8 4 2 Disinfecting 8 4 3 Maintenance 8 5 DISPOSING OF OBSOLETE EQUIPMENT 9 6 ELECTROMAGNETIC COMPATIBILITY 9 6 1 Electromagnetic Emissions 9 6 2 Electromagnetic Immunity 10 6 3 Recommended Separation Distances Between Portable and Mobile RF Communication Equipment and RiX 70DC 11 Appendix A Systems 12 Appendix B Icons 14 Appendix C Heating and Cooling Curves 15 Appendix D Alarms 16 A...

Страница 3: ... performances which could cause death injuries or health hazard to Patient and or Operator 1 2 Safety Recommendations Carefully read all the safety instructions to avoid potential hazards that could result in personal injuries or could damage your equipment This table contains important information regarding the system the operator and the patient safety Please read this table very carefully RiX 7...

Страница 4: ... operation of the system by involving a specialist for any maintenance Do not use the device in the presence of vapours or flammable anaesthetic mixtures with air or oxygen or nitrous oxide When using digital sensors if the PC is positioned within the patient environment it must meet the requirements defined by the IEC 60601 1 medical device if located outside it must be in compliance with IEC 609...

Страница 5: ...cm Beam diameter X Ray at 200 mm focus 60 mm Bean limiting device additional Rectangular 36 x45 mm Dose intensity accuracy 50 2 3 X Ray Tube Manufacturer Toshiba Japan CEI Italy Model D 045 OX70 4 Inherent filtration 1 mm Al eq 0 5 mm Al eq Target angle 12 5 16 Filament current Max 3 A 1 5 2 1 A Max filament current 3 V 1 7 3 1 V Anodic thermal capacity 4 3 kJ 7 kJ Max anodic dissipation 100 W 110...

Страница 6: ...u it s necessary to push again Menu button X RAY button Activation pressing this button will be activated on the display a count down during which you can make expositions by pressing the X Ray button Tooth Type selection choose the tooth to be examined lower incisors lower canines lower bicuspids lower molars upper molars bicuspids upper canines upper teeth bitewing System ready green light ON kV...

Страница 7: ... symbol of your selection in the case of analog receptor it is also indicated the sensitivity of the selected film 3 Select the patient size and type of tooth Tooth type touch the symbol corresponding to the tooth you want to x ray the display will show the symbol corresponding to the selection Once you select these parameters RiX 70 DC automatically selects the output time for the type of X Ray t...

Страница 8: ...ds on many aspects as per the user preference contrast brightness etc and per patient mouth anatomy 70 kV Tooth Patient size Digital sensor Film type D Film type E Film type F Phosphor system Adult Child Adult Child Adult Child Adult Child Adult Child Maxilla Molars 0 100 0 063 0 320 0 200 0 125 0 080 0 100 0 063 0 320 0 200 Premolars 0 080 0 050 0 250 0 160 0 100 0 063 0 080 0 050 0 250 0 160 Inc...

Страница 9: ...5 0 028 0 032 0 036 0 040 0 045 0 050 0 056 0 063 0 0710 0 080 0 090 0 100 0 110 0 125 0 140 0 160 0 180 0 200 0 220 0 250 0 280 0 320 0 360 0 400 0 450 0 500 0 560 0 630 0 710 0 800 0 900 1 000 1 100 1 250 1 400 1 600 1 800 2 000 4 CARE AND MAINTENANCE 4 1 Cleaning Always disconnect the line voltage supply before cleaning the unit Use a mild soap to remove fin ger or other dirty marks paying atte...

Страница 10: ...ribed below 6 1 Electromagnetic Emissions The device may cause radio interference and may disrupt the operation of nearby equipment It may be necessary to take mitigation measures such as re orienting or relocating the equipment or shielding the location The system comes with all the cables needed for its operation Using other cables not supplied by the manufacturer or in addition to those provide...

Страница 11: ...al or hospital environment If the user of the device requires continued operation during power mains interruptions the device needs to be powered from an uninterruptible power source Power frequency 50 60Hz magnetic field IEC 61000 4 8 3 A m Test level IEC 60601 1 2 Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environm...

Страница 12: ...distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be determined using the equation applicable to the frequency of the tra...

Страница 13: ...0 DC 12 Appendix A Systems 1 2 7 6 3 4 5 8 Walll mounted version 1 Wall plate 2 Support arm 40 60 80 cms 3 Scissor arm 4 Monoblock Support tie 5 Monoblock 6 Cone collimator 7 Control panel timer 8 Handswitch with cable ...

Страница 14: ...User Manual RiX 70 DC 13 Stand mobile version 1 Mobile stand base 2 Mobile stand column 3 Control unit timer hand switch with cable 4 Scissor arm 5 Monoblock with cone collimator 1 2 3 4 5 ...

Страница 15: ...ended function The product at the end of its useful life can not be thrown in the regular trash with other wastes and is subject to a separate collection Alternating Current N Conductor connected to the neutral point L Connection point to the Power Line Grounding Functional earth connection ON OFF POWER ON Power OFF Electric shock risk Product Reference Serial Number Manufacturing date Manufacture...

Страница 16: ...User Manual RiX 70 DC 15 Appendix C Heating and cooling curves CEI TUBE TOSHIBA TUBE Monoblock cooling curve ...

Страница 17: ...tion does not match the specifications Repeat the configuration of the system Call the tech assistance to replace the CPU board if the problem occurs frequently Error 1 X Ray button pressed No operating system Release the keys eventually pressed turn the system off and on again Consult the Maintenance Manual Error 5 More than one button pressed during the switching on System inoperative Release th...

Страница 18: ...rror 8 Open door The room s door if equipped with the proper sensor has been opened during the phase Ready for X RAY Close the door and restart the operations pressing Menu button Blocking messages These messages identify warnings situations for the patient the operator and other persons located close to the machine For these reason these errors block the machine Display info Description Cause Sol...

Страница 19: ... N Label 3 Scissor Double pantograph Arm Model it s a fixed code that identifies the double pantograph arm E00 03 Serial Number It s an alphanumeric code of 2 letters 6 numbers for instance EA000001 EA identifies the double pantograph arm 00001 identifies the sequence productive number Label 4 Collimator Bean Limiting Device REF E00 04 00 021 it s a fixed code that identifies the rectangular colli...

Страница 20: ...trident dental com info trident dental com IMPORTANT Please do not try to open your unit Only trained and qualified technicians authorized by Trident can remove the covers giving access to the circuits Please contact an authorized Trident distributor responsible for the technical assistance of your device ...