User Manual
RiX 70 DC
16
Appendix D
Alarms
Display info
Description
Cause / Solution
------
The display is completely black
No power. Reset and switch on again ; if the error
persists, please consult the Maintenance Manual
If the power switch and the indicator lights are switched
on, probably the main fuse is broken .
Consult the Maintenance Manual.
-------
The display is completely blank
even if the switch is on (green
light on)
Likely failure of the main fuse. Consult the Maintenance
Manual.
If the error persists, fatal error of the program memory.
Call the tech assistance to Replace the CPU board.
Error 9
Memory Error EEPROM 1
Control error of the first memory bank data ; the
system is operating as it has restored the data from
the second memory bank . Press " Menu " to continue
checking.
Call the tech assistance to replace the CPU board if
the problem occurs frequently
Error 9
Memory Error EEPROM 2
Control error of both the first and the second memory
bank; the system is operating as it has restored the
data from the program memory. Press " Menu " to
continue checking .
Call the tech assistance to replace the CPU board if
the problem occurs frequently
Error 10
Non congruent configuration
The system detected that the system configuration
does not match the specifications.
Repeat the configuration of the system.
Call the tech assistance to replace the CPU board if
the problem occurs frequently
Error 1
X-Ray button pressed
No operating system
Release the keys eventually pressed; turn the system
off and on again.
Consult the Maintenance Manual.
Error 5
More than one button pressed
during the switching on.
System inoperative.
Release the buttons; turn the system off. On again.
If the problem persists, please consult the
Maintenance Manual.
Error during pre-exposure phase
Display info
Description
Cause / Solution
Mains voltage too low.
The system supply voltage has dropped below the
minimum level allowed for the security of the system.
Press " Menu " to delete the message . If it reappears or
if the situation frequently happens, verify that the
office’s supply voltage is within the tolerances.
Error 1
Button
pressed
without
enabling
The system is protected by a hardware protection key
that was not correctly positioned before pressing the
X-Ray button .
Release the button, press the "Menu" button to resume
the system operation of and turn the security key to the
enabled position before repeating the exposure.
Release the button and press the "Menu" button to
resume operation of the system.
Enter the state of "Ready for X-Rays" before performing
the exposure.
In the remote system.
X-Ray button was not simultaneously pressed with "+"
button to enable the emission.