1    Introduction 

Thank you for trusting our company and choosing RiX-70 DC as your Intraoral X-Ray System.   
 
This manual, supplied with the system, is an integral part of the product; please keep it close to the 
unit’s installation location; it describes the parts and components of your RiX-70 DC and explains the 
correct  operation  of the unit by  observing all the relevant  safety  measures when handling X-ray 
equipment. The original language of the User Manual is Italian. 
 

1.1 User Recommendations 

 

 

Carefully read and follow the warnings and safety instructions

 

contained in the 

User’s Manual.

 

  

 

Use  the  device  in  accordance  with  the  procedures  explained  in  the  present 
manual,  and do not  use it for  purposes  other than those it was  designed for. 
Failure of the user to properly maintain the equipment may relieve Trident or his 
dealer,  from  responsibility  for  any  injury,  damage,  or  non-compliance,  which 
may result. 

 

Report promptly to  Trident or to its dealer any accident involving this device or 
any  alteration  in  features  and/or  performances  which  could  cause  death, 
injuries or health hazard to Patient and/or Operator. 

 
 

1.2  Safety Recommendations  

Carefully  read  all  the  safety  instructions  to  avoid  potential  hazards  that  could  result  in  personal 
injuries or could damage your equipment. 
 

 This  table  contains  important  information  regarding  the  system,  the  operator 
and the patient safety. 
Please read this table very carefully. 

 

 

RiX-70 DC is an electro-medical device that must be used under the supervision 
of a doctor or a person with vast experience in radiology. 

 

Do not modify the device mechanically and / or electrically during or after the 
installation.   

 

RiX-70 DC is built for continuous operation with intermittent load; therefore, it is 
required  to  observe  the  cycles  of  use  allowing  the  dissipation  of  heat 
accumulated by the radiation source. 

 

RiX-70 DC must be switched off when using electro surgery or similar devices. 

 

Make  sure  the  installation  of  your  X-Ray  unit  is  carried  out    only  by  qualified 
personnel.  Absolutely  avoid  provisional  electrical  connections  and  verify  that 
only the parts (screws, nuts, bolts, etc.) recommended by the manufacturer are 
used. 

 

The  installation  area  must  comply  with  the  current  regulations  pertaining  to 
radiology in the country of operation. 

Содержание RiX-70 DC

Страница 1: ...User Manual RiX 70 DC i User Manual RiX 70 DC X Ray Intraoral System Date 19 01 2017 Language English Rev 2 Code UM04 EN 0051 ...

Страница 2: ... 4 1 Cleaning 8 4 2 Disinfecting 8 4 3 Maintenance 8 5 DISPOSING OF OBSOLETE EQUIPMENT 9 6 ELECTROMAGNETIC COMPATIBILITY 9 6 1 Electromagnetic Emissions 9 6 2 Electromagnetic Immunity 10 6 3 Recommended Separation Distances Between Portable and Mobile RF Communication Equipment and RiX 70DC 11 Appendix A Systems 12 Appendix B Icons 14 Appendix C Heating and Cooling Curves 15 Appendix D Alarms 16 A...

Страница 3: ... performances which could cause death injuries or health hazard to Patient and or Operator 1 2 Safety Recommendations Carefully read all the safety instructions to avoid potential hazards that could result in personal injuries or could damage your equipment This table contains important information regarding the system the operator and the patient safety Please read this table very carefully RiX 7...

Страница 4: ... operation of the system by involving a specialist for any maintenance Do not use the device in the presence of vapours or flammable anaesthetic mixtures with air or oxygen or nitrous oxide When using digital sensors if the PC is positioned within the patient environment it must meet the requirements defined by the IEC 60601 1 medical device if located outside it must be in compliance with IEC 609...

Страница 5: ...cm Beam diameter X Ray at 200 mm focus 60 mm Bean limiting device additional Rectangular 36 x45 mm Dose intensity accuracy 50 2 3 X Ray Tube Manufacturer Toshiba Japan CEI Italy Model D 045 OX70 4 Inherent filtration 1 mm Al eq 0 5 mm Al eq Target angle 12 5 16 Filament current Max 3 A 1 5 2 1 A Max filament current 3 V 1 7 3 1 V Anodic thermal capacity 4 3 kJ 7 kJ Max anodic dissipation 100 W 110...

Страница 6: ...u it s necessary to push again Menu button X RAY button Activation pressing this button will be activated on the display a count down during which you can make expositions by pressing the X Ray button Tooth Type selection choose the tooth to be examined lower incisors lower canines lower bicuspids lower molars upper molars bicuspids upper canines upper teeth bitewing System ready green light ON kV...

Страница 7: ... symbol of your selection in the case of analog receptor it is also indicated the sensitivity of the selected film 3 Select the patient size and type of tooth Tooth type touch the symbol corresponding to the tooth you want to x ray the display will show the symbol corresponding to the selection Once you select these parameters RiX 70 DC automatically selects the output time for the type of X Ray t...

Страница 8: ...ds on many aspects as per the user preference contrast brightness etc and per patient mouth anatomy 70 kV Tooth Patient size Digital sensor Film type D Film type E Film type F Phosphor system Adult Child Adult Child Adult Child Adult Child Adult Child Maxilla Molars 0 100 0 063 0 320 0 200 0 125 0 080 0 100 0 063 0 320 0 200 Premolars 0 080 0 050 0 250 0 160 0 100 0 063 0 080 0 050 0 250 0 160 Inc...

Страница 9: ...5 0 028 0 032 0 036 0 040 0 045 0 050 0 056 0 063 0 0710 0 080 0 090 0 100 0 110 0 125 0 140 0 160 0 180 0 200 0 220 0 250 0 280 0 320 0 360 0 400 0 450 0 500 0 560 0 630 0 710 0 800 0 900 1 000 1 100 1 250 1 400 1 600 1 800 2 000 4 CARE AND MAINTENANCE 4 1 Cleaning Always disconnect the line voltage supply before cleaning the unit Use a mild soap to remove fin ger or other dirty marks paying atte...

Страница 10: ...ribed below 6 1 Electromagnetic Emissions The device may cause radio interference and may disrupt the operation of nearby equipment It may be necessary to take mitigation measures such as re orienting or relocating the equipment or shielding the location The system comes with all the cables needed for its operation Using other cables not supplied by the manufacturer or in addition to those provide...

Страница 11: ...al or hospital environment If the user of the device requires continued operation during power mains interruptions the device needs to be powered from an uninterruptible power source Power frequency 50 60Hz magnetic field IEC 61000 4 8 3 A m Test level IEC 60601 1 2 Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environm...

Страница 12: ...distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be determined using the equation applicable to the frequency of the tra...

Страница 13: ...0 DC 12 Appendix A Systems 1 2 7 6 3 4 5 8 Walll mounted version 1 Wall plate 2 Support arm 40 60 80 cms 3 Scissor arm 4 Monoblock Support tie 5 Monoblock 6 Cone collimator 7 Control panel timer 8 Handswitch with cable ...

Страница 14: ...User Manual RiX 70 DC 13 Stand mobile version 1 Mobile stand base 2 Mobile stand column 3 Control unit timer hand switch with cable 4 Scissor arm 5 Monoblock with cone collimator 1 2 3 4 5 ...

Страница 15: ...ended function The product at the end of its useful life can not be thrown in the regular trash with other wastes and is subject to a separate collection Alternating Current N Conductor connected to the neutral point L Connection point to the Power Line Grounding Functional earth connection ON OFF POWER ON Power OFF Electric shock risk Product Reference Serial Number Manufacturing date Manufacture...

Страница 16: ...User Manual RiX 70 DC 15 Appendix C Heating and cooling curves CEI TUBE TOSHIBA TUBE Monoblock cooling curve ...

Страница 17: ...tion does not match the specifications Repeat the configuration of the system Call the tech assistance to replace the CPU board if the problem occurs frequently Error 1 X Ray button pressed No operating system Release the keys eventually pressed turn the system off and on again Consult the Maintenance Manual Error 5 More than one button pressed during the switching on System inoperative Release th...

Страница 18: ...rror 8 Open door The room s door if equipped with the proper sensor has been opened during the phase Ready for X RAY Close the door and restart the operations pressing Menu button Blocking messages These messages identify warnings situations for the patient the operator and other persons located close to the machine For these reason these errors block the machine Display info Description Cause Sol...

Страница 19: ... N Label 3 Scissor Double pantograph Arm Model it s a fixed code that identifies the double pantograph arm E00 03 Serial Number It s an alphanumeric code of 2 letters 6 numbers for instance EA000001 EA identifies the double pantograph arm 00001 identifies the sequence productive number Label 4 Collimator Bean Limiting Device REF E00 04 00 021 it s a fixed code that identifies the rectangular colli...

Страница 20: ...trident dental com info trident dental com IMPORTANT Please do not try to open your unit Only trained and qualified technicians authorized by Trident can remove the covers giving access to the circuits Please contact an authorized Trident distributor responsible for the technical assistance of your device ...