User Manual
RiX 70 DC
17
Display info
Description
Cause / Solution
Error 1
X RAY Button pressed after the
end of the cooling break
Release the X RAY button and press “Menu” button. If
the condition persists , there is a probable damage in
the X RAY button. To see more information please refer
to the Maintenance Manual.
Error 7
Communication error RS485
Caution: blocking error.
Communication between system boards does not work
correctly. Turn the system off and on again; If the error
persists, please consult the Maintenance Manual.
Error during X RAY emission
Display info
Description
Cause / Solution
Error 2
Exposure ended by the operator
Releasing of X-Ray button during the emission has
interrupted the emission.
In case of phosphor or analogic receptor:
change the film and repeat the emission.
In case of digital sensor: The image obtained
can be not diagnostic, repeat the operations.
Error 1
X-RAY button pressed
The X RAY button is already pressed and,
accidentally, another button is pressed.
If the error message doesn’t disappear, even after
switching off and on the system, please consult
the Maintenance Manual.
Error 8
Open door
The room’s door, if equipped with the proper
sensor, has been opened during the phase
“Ready for X RAY”. Close the door and restart the
operations pressing “Menu” button.
Blocking messages
These messages identify warnings situations for the patient, the operator and other persons located close
to the machine. For these reason these errors block the machine.
Display info
Description
Cause / Solution
Error 3
End of exposure by security timer
The emission did not end normally and was
interrupted by the safety timer . The audio signal
will be active until the intervention of the security .
The image is not diagnostic.
The system is blocked: switch off and on again .
If the problem persists please call tech assistance
to replace the driver board.
Error 6
Overvoltage security intervention
The system has detected that high voltage values
are too high; the emission was interrupted.
The image is not diagnostic.
The system is blocked: switch off and on again.
If the problem persists, call tech assistance to
replace the driver board.
Error 4
Error filament power supply
The system detected the lack of the anode current
feedback signal.
The emission cannot start or was cancelled.
The image is not diagnostic.
The system is locked: turn off and on again.
If the problem persists, please consult the
Maintenance Manual.