Trident RiX-70 DC Скачать руководство пользователя страница 4

 
 

 

User Manual 

RiX 70 DC

 

 

 

 

Do  not  drop,  knock,  or  shake  the  unit:  rough  handling  or  use  beyond  the 
recommendations may cause damage to internal circuitry. 

 

The operator should take every reasonable precaution to maintain a safe and 
healthy working environment,   minimizing the radiation hazard. 

 

During the exposure, make sure no one except the patient and operator are 
present in the X-ray area.

 

 

In accordance with EN 60601-1, it is strictly forbidden to modify the equipment 
or its parts. 
Only  trained  and  qualified  technicians  are  authorized  to  remove  the  covers 
giving access to the circuits.  
The cables of the power supply must comply with applicable laws and must be 
equipped with terminals for connecting to the protective earth. 

 

Always turn off the machine and unplug it from the mains whenever possible 
(with the switch) before cleaning and disinfection. 

 

Carry  out  a  proper  cleaning  and  disinfection  of  all  parts  which  come  into 
contact with the patient. 

 

Check  regularly  (at  least  once  a  year)  the  state  of  correct  operation  of  the 
system by involving a specialist for any maintenance. 

 

Do not use the device in the presence  of vapours, or flammable anaesthetic 
mixtures with air, or oxygen or nitrous oxide. 

 

When using digital sensors, if the PC is positioned within the patient environment, 
it  must  meet  the  requirements  defined  by  the  IEC  60601-1  medical  device;  if 
located outside it must be in compliance with IEC 60950. The size of the patient 
environment is defined as a minimum distance of 1.5 m from the patient. 

 

1.3  About your unit 

RiX-70 DC is an intraoral X-ray system that emits ionizing radiation,  X Ray, for taking images of the 
oral cavity. RiX-70 DC with a high technological development allows getting high quality images 
using both normal films and digital images receptors (intraoral sensor and phosphor plates).  This 
system has two version, wall mounted and mobile. 
 

1.4  Device Classification 

RiX-70 DC is a Class I device, Type B, conforms to IEC 60601-2-65: 2012.  RiX-70 DC, according to the 
classification rules set out in Annex IX of EC Directive 93/42 and its amendments and supplements,   
is  an  active  medical  device  that  releases  energy  (ionizing  radiation  X  Ray)  and  represents  a 
potential hazard to the human body.    
 
 

 

Содержание RiX-70 DC

Страница 1: ...User Manual RiX 70 DC i User Manual RiX 70 DC X Ray Intraoral System Date 19 01 2017 Language English Rev 2 Code UM04 EN 0051 ...

Страница 2: ... 4 1 Cleaning 8 4 2 Disinfecting 8 4 3 Maintenance 8 5 DISPOSING OF OBSOLETE EQUIPMENT 9 6 ELECTROMAGNETIC COMPATIBILITY 9 6 1 Electromagnetic Emissions 9 6 2 Electromagnetic Immunity 10 6 3 Recommended Separation Distances Between Portable and Mobile RF Communication Equipment and RiX 70DC 11 Appendix A Systems 12 Appendix B Icons 14 Appendix C Heating and Cooling Curves 15 Appendix D Alarms 16 A...

Страница 3: ... performances which could cause death injuries or health hazard to Patient and or Operator 1 2 Safety Recommendations Carefully read all the safety instructions to avoid potential hazards that could result in personal injuries or could damage your equipment This table contains important information regarding the system the operator and the patient safety Please read this table very carefully RiX 7...

Страница 4: ... operation of the system by involving a specialist for any maintenance Do not use the device in the presence of vapours or flammable anaesthetic mixtures with air or oxygen or nitrous oxide When using digital sensors if the PC is positioned within the patient environment it must meet the requirements defined by the IEC 60601 1 medical device if located outside it must be in compliance with IEC 609...

Страница 5: ...cm Beam diameter X Ray at 200 mm focus 60 mm Bean limiting device additional Rectangular 36 x45 mm Dose intensity accuracy 50 2 3 X Ray Tube Manufacturer Toshiba Japan CEI Italy Model D 045 OX70 4 Inherent filtration 1 mm Al eq 0 5 mm Al eq Target angle 12 5 16 Filament current Max 3 A 1 5 2 1 A Max filament current 3 V 1 7 3 1 V Anodic thermal capacity 4 3 kJ 7 kJ Max anodic dissipation 100 W 110...

Страница 6: ...u it s necessary to push again Menu button X RAY button Activation pressing this button will be activated on the display a count down during which you can make expositions by pressing the X Ray button Tooth Type selection choose the tooth to be examined lower incisors lower canines lower bicuspids lower molars upper molars bicuspids upper canines upper teeth bitewing System ready green light ON kV...

Страница 7: ... symbol of your selection in the case of analog receptor it is also indicated the sensitivity of the selected film 3 Select the patient size and type of tooth Tooth type touch the symbol corresponding to the tooth you want to x ray the display will show the symbol corresponding to the selection Once you select these parameters RiX 70 DC automatically selects the output time for the type of X Ray t...

Страница 8: ...ds on many aspects as per the user preference contrast brightness etc and per patient mouth anatomy 70 kV Tooth Patient size Digital sensor Film type D Film type E Film type F Phosphor system Adult Child Adult Child Adult Child Adult Child Adult Child Maxilla Molars 0 100 0 063 0 320 0 200 0 125 0 080 0 100 0 063 0 320 0 200 Premolars 0 080 0 050 0 250 0 160 0 100 0 063 0 080 0 050 0 250 0 160 Inc...

Страница 9: ...5 0 028 0 032 0 036 0 040 0 045 0 050 0 056 0 063 0 0710 0 080 0 090 0 100 0 110 0 125 0 140 0 160 0 180 0 200 0 220 0 250 0 280 0 320 0 360 0 400 0 450 0 500 0 560 0 630 0 710 0 800 0 900 1 000 1 100 1 250 1 400 1 600 1 800 2 000 4 CARE AND MAINTENANCE 4 1 Cleaning Always disconnect the line voltage supply before cleaning the unit Use a mild soap to remove fin ger or other dirty marks paying atte...

Страница 10: ...ribed below 6 1 Electromagnetic Emissions The device may cause radio interference and may disrupt the operation of nearby equipment It may be necessary to take mitigation measures such as re orienting or relocating the equipment or shielding the location The system comes with all the cables needed for its operation Using other cables not supplied by the manufacturer or in addition to those provide...

Страница 11: ...al or hospital environment If the user of the device requires continued operation during power mains interruptions the device needs to be powered from an uninterruptible power source Power frequency 50 60Hz magnetic field IEC 61000 4 8 3 A m Test level IEC 60601 1 2 Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environm...

Страница 12: ...distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be determined using the equation applicable to the frequency of the tra...

Страница 13: ...0 DC 12 Appendix A Systems 1 2 7 6 3 4 5 8 Walll mounted version 1 Wall plate 2 Support arm 40 60 80 cms 3 Scissor arm 4 Monoblock Support tie 5 Monoblock 6 Cone collimator 7 Control panel timer 8 Handswitch with cable ...

Страница 14: ...User Manual RiX 70 DC 13 Stand mobile version 1 Mobile stand base 2 Mobile stand column 3 Control unit timer hand switch with cable 4 Scissor arm 5 Monoblock with cone collimator 1 2 3 4 5 ...

Страница 15: ...ended function The product at the end of its useful life can not be thrown in the regular trash with other wastes and is subject to a separate collection Alternating Current N Conductor connected to the neutral point L Connection point to the Power Line Grounding Functional earth connection ON OFF POWER ON Power OFF Electric shock risk Product Reference Serial Number Manufacturing date Manufacture...

Страница 16: ...User Manual RiX 70 DC 15 Appendix C Heating and cooling curves CEI TUBE TOSHIBA TUBE Monoblock cooling curve ...

Страница 17: ...tion does not match the specifications Repeat the configuration of the system Call the tech assistance to replace the CPU board if the problem occurs frequently Error 1 X Ray button pressed No operating system Release the keys eventually pressed turn the system off and on again Consult the Maintenance Manual Error 5 More than one button pressed during the switching on System inoperative Release th...

Страница 18: ...rror 8 Open door The room s door if equipped with the proper sensor has been opened during the phase Ready for X RAY Close the door and restart the operations pressing Menu button Blocking messages These messages identify warnings situations for the patient the operator and other persons located close to the machine For these reason these errors block the machine Display info Description Cause Sol...

Страница 19: ... N Label 3 Scissor Double pantograph Arm Model it s a fixed code that identifies the double pantograph arm E00 03 Serial Number It s an alphanumeric code of 2 letters 6 numbers for instance EA000001 EA identifies the double pantograph arm 00001 identifies the sequence productive number Label 4 Collimator Bean Limiting Device REF E00 04 00 021 it s a fixed code that identifies the rectangular colli...

Страница 20: ...trident dental com info trident dental com IMPORTANT Please do not try to open your unit Only trained and qualified technicians authorized by Trident can remove the covers giving access to the circuits Please contact an authorized Trident distributor responsible for the technical assistance of your device ...

Отзывы: