VMS-1655M
LIT-12013297
21
Optional flow sensor setup
The controller can accept three input signal ranges (0 V to 5 V, 0 V to 10 V, 4 mA to 20 mA) from a pressure or flow sensor that
attaches to a flow probe in the duct work. This signal computes a cfm value that, with scaling, outputs on the 4 mA to 20 mA output
signal. Information and tools to set this up are available in the
CFM Settings
tab.
Figure 28: CFM settings
The table in the tool updates with values that correspond to the high and low face velocities, as well as the max face velocity.
Figure 27: Updated Hood Charactization table