background image

The TMB-

1014-BT

is intended for use in the electromagnetic environment specified

below. The customer or the user of TMB-

1014-BT

should assure that it is used in

such an environment.

Guidance and manufacture’s declaration – electromagnetic immunity

Immunity test

±6 kV contact
±8 kV air

±6 kV contact
±8 kV air

±2 kV for
power supply lines

±2kV for
power supply lines

±1 kV line(s)
to line(s)

±1 kV
differential mode

<5% UT
(>95% dip in UT)
for 0.5 cycle

40% UT
(60% dip in UT)
for 5 cycles

70% UT
(30% dip in UT)
for 25 cycles

<5% UT
(>95% dip in UT)
for 5 sec

<5% UT
(>95% dip in UT)
for 0.5 cycle

40% UT)
(60% dip in UT)
for 5 cycles

70% UT
(30% dip in UT)
for 25 cycles

<5% UT
(>95% dip in UT)
for 5 sec

3A/m

NOTE

UT is the a.c. mains voltage prior to application of the test level.

Table 2 Guidance and manufacture’s declaration – electromagnetic immunity –

for all ME EQUIPMENT and ME SYSTEMS

IEC 60601 test level Compliance level

Electrostatic
discharge (ESD)
IEC 61000-4-2

Electromagnetic
environment - guidance

Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.

Mains power quality should
be that of a typical
commercial or hospital
environment.

Mains power quality should
be that of a typical
commercial or hospital
environment.

Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the TMB-

1014-BT

requires

continued operation during
power mains interruptions,
it is recommended that the
TMB-

1014-BT

be powered

from an uninterruptible
power supply or a battery.

Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.

3A/m

Power frequency
(50Hz) magnetic
field IEC
61000-4-8

Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11

Electrical fast
transient/burst
IEC 61000-4-4

Surge IEC
61000-4-5

Table 4 Guidance and manufacture’s declaration – electromagnetic immunity –

for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Guidance and manufacture’s declaration – electromagnetic immunity

Immunity test

3 Vrms

3 V/m

The TMB-

1014-BT

is intended for use in the electromagnetic environment specified

below. The customer or the user should assure that it is used in such an environment.

Compliance
level

IEC 60601
test level

Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

3 V/m 80 MHz
to 2.5 GHz

3 V

rms

150 kHz to
80 MHz

Electromagnetic environment -
guidance

Portable and mobile RF communications
equipment should be used no closer to
any part of the YS-6100, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Portable and mobile RF communications
equipment should be used no closer to
any part of the including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.

Recommended separation distance

d = 1.167
d = 1.167

80 MHz to 800 MHz

d = 2.333

800 MHz to 2.5 GHz

where

P

is the maximum output power

rating of the transmitter in watts (W)
according to the transmitter manufac-
turer and

d

 is the recommended

separation distance in metres (m).

Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,

a

should be

less than the compliance level in each
frequency range.

Interference may occur in the vicinity of
equipment marked with the following
symbol:

Содержание TMB-1014-BT

Страница 1: ...please read the manual thoroughly Thank you very much for selecting TRANSTEK Blood Pressure Monitor TMB 1014 BT Please well keep this manual in order to reference in future Zhongshan Transtek Electron...

Страница 2: ...matically deflate Should the cuff not deflate when pressures exceeds 40 kPa 300 mmHg detach the cuff from the wrist and press the START STOP button to stop inflation To avoid measurement errors carefu...

Страница 3: ...gh pressure result Diastolic blood pressure Low pressure result Pulse Pulse minute mmHg Measurement unit the blood pressure Time Currently time Low battery Batteries are low and need to be replaced Er...

Страница 4: ...mory In sure the lock key is glided to the unlock mode year 2010 2050 time 24 H Remove batteries if the device is not likely to be used for some time The old battery is harmful to the environment so p...

Страница 5: ...column will remind searching for device 3 Long press the START button after power off the blood pressure monitor it ll show moving o 4 At this moment the discovered device column of mobile will displ...

Страница 6: ...blood pressure monitor show it means matching successfully When it display E1 it indicates matching unsuccessfully 7 If matching successfully it will show the lately measuring record on the receive d...

Страница 7: ...atically Display and save the results automatically Start the Measurement 4 Press the START STOP to power off otherwise it will turn off automatically within 1 minute 2 After measuring it ll display t...

Страница 8: ...3 and so on and the last record 60 is dropped from the list CAUTION UP DOWN MEM SET MEM SET If you did not get the correct measurement you can delete all results by following below steps Deleting a me...

Страница 9: ...lision Using the wet clothing to remove the dirt Avoid touching water clean it with a dry cloth in case Avoid the dusty and unstable temperature environment Avoid washing the cuff Clean the cuff with...

Страница 10: ...too loose If the cuff is tied on the wrist If you feel anxious pressured You had better take deep breath 2 3 times before beginning Advice adjust yourself for 4 5 minutes until you calm down ABOUT BLO...

Страница 11: ...the products not operating as you think it should check here before arranging for servicing PROBLEM SYMPTOM CHECK THIS REMEDY No power Low batteries Error massage Display is dim or will not light up...

Страница 12: ...ission CISPR 11 The TMB 1014 BT uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment The TM...

Страница 13: ...location in a typical commercial or hospital environment 3A m Power frequency 50Hz magnetic field IEC 61000 4 8 Voltage dips short interruptions and voltage variations on power supply input lines IEC...

Страница 14: ...equipment and the EQUIPMENT or SYSTEM for ME EQUIPMENT or ME SYSTEM that are not LIFE SUPPORTING Recommended separation distances between portable and mobile RF communica tions equipment and the ELE00...

Страница 15: ...and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this...