The TMB-
1014-BT
is intended for use in the electromagnetic environment specified
below. The customer or the user of TMB-
1014-BT
should assure that it is used in
such an environment.
Guidance and manufacture’s declaration – electromagnetic immunity
Immunity test
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
±2 kV for
power supply lines
±2kV for
power supply lines
±1 kV line(s)
to line(s)
±1 kV
differential mode
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT)
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
3A/m
NOTE
UT is the a.c. mains voltage prior to application of the test level.
Table 2 Guidance and manufacture’s declaration – electromagnetic immunity –
for all ME EQUIPMENT and ME SYSTEMS
IEC 60601 test level Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the TMB-
1014-BT
requires
continued operation during
power mains interruptions,
it is recommended that the
TMB-
1014-BT
be powered
from an uninterruptible
power supply or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
3A/m
Power frequency
(50Hz) magnetic
field IEC
61000-4-8
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
Electrical fast
transient/burst
IEC 61000-4-4
Surge IEC
61000-4-5
Table 4 Guidance and manufacture’s declaration – electromagnetic immunity –
for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
Immunity test
3 Vrms
3 V/m
The TMB-
1014-BT
is intended for use in the electromagnetic environment specified
below. The customer or the user should assure that it is used in such an environment.
Compliance
level
IEC 60601
test level
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz
to 2.5 GHz
3 V
rms
150 kHz to
80 MHz
Electromagnetic environment -
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the YS-6100, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Portable and mobile RF communications
equipment should be used no closer to
any part of the including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d = 1.167
d = 1.167
80 MHz to 800 MHz
d = 2.333
800 MHz to 2.5 GHz
where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufac-
turer and
d
is the recommended
separation distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol: