Specifications and displayed error codes.

General

Model

Capacity /
Graduation

Weight kg

Dimension mm

Battery

Weighblok

Max 100 kg / 50 g
Not approved version

4,5

260 x 345 x 55

6 of alk. 1,5 V batteries LR6/AA for 15 hours operation.

OPERATOR INTERFACE

Display

0.75" (19 mm) 7-segment, Liquid Crystal, 6-Digit

Additional Symbols

Net, Gross, Stable, Tare, lb, kg, Zero

Keyboard

5-key flat membrane panel

ENVIRONMENTAL

Operating Temperature
Storage Temperature

Displayed error codes.

CODE

MODE

MEANING / POSSIBLE SOLUTION

rrrrrr

Normal Operating
Mode

Gross Overload. A weight greater than the rated capacity has been applied to the scale. Remove the weight
from the platter or try re-calibrating the scale. Otherwise, check for a bad load cell connection or possible
load cell damage due to overloading.

bAtt

Normal Operating
Mode

Indicates a low battery condition. Re-charge the battery or replace if alkaline batteries.

Err 0

Span Calibration
Mode (F17)

Keyed-in weight value is larger than full-scale capacity. Use a smaller test weight or check keyed-in value.

Err 1

Span Calibration
Mode (F17)

Keyed-in weight value is less than 1% of full-scale capacity. Use a larger test weight or check keyed-in
value.

Err 2

Span Calibration
Mode (F17)

There is not enough load cell signal to produce the internal counts necessary to properly calibrate the scale.
First check all load connections. Use F16 mode to view internal counts. See Appendix C for more
information.

Err 3

All Modes

Non-volatile memory read error. One or more setup parameters have been lost.

Err 4

All Modes

Non-volatile memory writes error. Indicator needs service.

Err 5

Key-in Span
Calibration Mode
(F20)

You have attempted to enter a zero value for C1. Enter a known calibration value greater than zero.

Err 7

Initialization

No reading from the ADC. Make sure there is a load cell(s) connected to the indicator at start-up.

Err 9

Normal Operating
Mode

Span calibration value has been lost. Re-calibrate the Scale.

Содержание WEIGHBLOK

Страница 1: ...Revision 0502 4 2 03 2001 2002 Transcell Technology Inc Contents subject to change without notice Transcell Technology Inc 975 Deerfield Parkway Buffalo Grove IL 60089 Tel 847 419 9180 Fax 847 419 15...

Страница 2: ...Manual Normal Calibration Calibration switch ON PRINT NET GROSS OFF ZERO TARE Communica tions port External power 9 VDC Alkaline batteries The Weighingblock is a robust Scale It is designed for portab...

Страница 3: ...enu while a few other parameters are altered through the User menu The configuration section of the manual explains how to use the five front panel keys to maneuver and save settings in both menus The...

Страница 4: ...ve to the selection value press ZERO once 7 Increase and decrease the value with PRINT and TARE 8 To save the value press NET 9 Press DOT key to go back to parameter number 10 To go back to normal men...

Страница 5: ...accuracy The higher the filter number the greater the accuracy but the slower the response time Choose 4 or 8 unless a very fast response is needed 1 2 4 8 F7 Overload Limit Selects the desired formul...

Страница 6: ...field Parkway 800 503 9180 www transcell net Buffalo Grove IL 60089 847 419 1515 transcell transcell net F6 F7 F8 F9 A10 Auto Power Off Period Selects the auto power off time period in minutes Off Dis...

Страница 7: ...omplete indicator calibration To adjust the max point using the F17 span adjusts procedure 1 Scroll to F 17 then press ZERO to enter span calibration menu 2 The display will momentarily show C 1 for t...

Страница 8: ...erating Mode Indicates a low battery condition Re charge the battery or replace if alkaline batteries Err 0 Span Calibration Mode F17 Keyed in weight value is larger than full scale capacity Use a sma...

Страница 9: ...normal mode Loadcell Connections Check the Loadcell with a universal voltmeter see table The left pin is number 1 Terminal J1 Description Colours Loadcell AG Colours Loadcell 1042 E 1 Out 5 VDC Brown...

Страница 10: ...appears Basic scale use 1 Press ON to start the scale 2 If necessary press ZERO to obtain a weight reading of zero 3 Press OFF to switch off the scale Hold 3 seconds 4 The scale switches off automatic...

Страница 11: ...when filling or discharging One time set up F11 must be 1 and A5 must be 2 normally done when delivered RS 232 PIN NUMBER PIN NAME 2 RXD 3 TXD 5 GND 7 TARGET TTL 5V OV 8 TTL 5V Set the target weight...

Отзывы: