Specifications and displayed error codes.
General
Model
Capacity /
Graduation
Weight kg
Dimension mm
Battery
Weighblok
Max 100 kg / 50 g
Not approved version
4,5
260 x 345 x 55
6 of alk. 1,5 V batteries LR6/AA for 15 hours operation.
OPERATOR INTERFACE
Display
0.75" (19 mm) 7-segment, Liquid Crystal, 6-Digit
Additional Symbols
Net, Gross, Stable, Tare, lb, kg, Zero
Keyboard
5-key flat membrane panel
ENVIRONMENTAL
Operating Temperature
Storage Temperature
Displayed error codes.
CODE
MODE
MEANING / POSSIBLE SOLUTION
rrrrrr
Normal Operating
Mode
Gross Overload. A weight greater than the rated capacity has been applied to the scale. Remove the weight
from the platter or try re-calibrating the scale. Otherwise, check for a bad load cell connection or possible
load cell damage due to overloading.
bAtt
Normal Operating
Mode
Indicates a low battery condition. Re-charge the battery or replace if alkaline batteries.
Err 0
Span Calibration
Mode (F17)
Keyed-in weight value is larger than full-scale capacity. Use a smaller test weight or check keyed-in value.
Err 1
Span Calibration
Mode (F17)
Keyed-in weight value is less than 1% of full-scale capacity. Use a larger test weight or check keyed-in
value.
Err 2
Span Calibration
Mode (F17)
There is not enough load cell signal to produce the internal counts necessary to properly calibrate the scale.
First check all load connections. Use F16 mode to view internal counts. See Appendix C for more
information.
Err 3
All Modes
Non-volatile memory read error. One or more setup parameters have been lost.
Err 4
All Modes
Non-volatile memory writes error. Indicator needs service.
Err 5
Key-in Span
Calibration Mode
(F20)
You have attempted to enter a zero value for C1. Enter a known calibration value greater than zero.
Err 7
Initialization
No reading from the ADC. Make sure there is a load cell(s) connected to the indicator at start-up.
Err 9
Normal Operating
Mode
Span calibration value has been lost. Re-calibrate the Scale.