Troubleshooting

If the Weighblok to power up.

1.

Change to new alkaline batteries.

2.

Check to see if AC adapter is plugged in.

If the scale starts but the display shows wrong.

1.

Adjust and Calibrate the Scale.

Bottom plate removal

1.

Remove the two screws from the bottom plate of the Weighblok.. Be careful with the keyboard connection.

2.

Check the inside of the Scale visually. Look for bad cables and connections.

If the Scale shows F1

1.

Check the position of the Calibration Switch.

2.

Reset to normal mode.

Loadcell Connections

Check the Loadcell with a universal voltmeter, see table. The left pin is number 1.

Terminal J1

Description

Colours Loadcell

AG

Colours Loadcell 1042

E+ (1)

Out +5 VDC*

Brown

Green

S+ (2)

In +

Yellow

Red

E- (3)

Out 0 VDC

Green

Black

S- (4)

In -

White

White

1.

Check the excitation voltage (E+ and E-) from the Indicator. Should be 5,0* (+-0,4) VDC.

2.

Check the input to the Indicator (S+ and S-). The output from the Loadcell S+ and S- (pin 2 and 4)
will increase from aprox. 0 to 10 mV analogue to the capacity range (100 or 30 kg). Note S+ is plus
and S- is minus. The polarity is important.

3.

If the input is wrong, <0 mV or > +10 mV with unloaded Scale check the Loadcell. Do have only
E+ and E- connected (or probably an external Power 5 VDC) and checks the output from the
Loadcells direct on the wires. Should be 0 mV to +10 mV depending on the load.

Содержание WEIGHBLOK

Страница 1: ...Revision 0502 4 2 03 2001 2002 Transcell Technology Inc Contents subject to change without notice Transcell Technology Inc 975 Deerfield Parkway Buffalo Grove IL 60089 Tel 847 419 9180 Fax 847 419 15...

Страница 2: ...Manual Normal Calibration Calibration switch ON PRINT NET GROSS OFF ZERO TARE Communica tions port External power 9 VDC Alkaline batteries The Weighingblock is a robust Scale It is designed for portab...

Страница 3: ...enu while a few other parameters are altered through the User menu The configuration section of the manual explains how to use the five front panel keys to maneuver and save settings in both menus The...

Страница 4: ...ve to the selection value press ZERO once 7 Increase and decrease the value with PRINT and TARE 8 To save the value press NET 9 Press DOT key to go back to parameter number 10 To go back to normal men...

Страница 5: ...accuracy The higher the filter number the greater the accuracy but the slower the response time Choose 4 or 8 unless a very fast response is needed 1 2 4 8 F7 Overload Limit Selects the desired formul...

Страница 6: ...field Parkway 800 503 9180 www transcell net Buffalo Grove IL 60089 847 419 1515 transcell transcell net F6 F7 F8 F9 A10 Auto Power Off Period Selects the auto power off time period in minutes Off Dis...

Страница 7: ...omplete indicator calibration To adjust the max point using the F17 span adjusts procedure 1 Scroll to F 17 then press ZERO to enter span calibration menu 2 The display will momentarily show C 1 for t...

Страница 8: ...erating Mode Indicates a low battery condition Re charge the battery or replace if alkaline batteries Err 0 Span Calibration Mode F17 Keyed in weight value is larger than full scale capacity Use a sma...

Страница 9: ...normal mode Loadcell Connections Check the Loadcell with a universal voltmeter see table The left pin is number 1 Terminal J1 Description Colours Loadcell AG Colours Loadcell 1042 E 1 Out 5 VDC Brown...

Страница 10: ...appears Basic scale use 1 Press ON to start the scale 2 If necessary press ZERO to obtain a weight reading of zero 3 Press OFF to switch off the scale Hold 3 seconds 4 The scale switches off automatic...

Страница 11: ...when filling or discharging One time set up F11 must be 1 and A5 must be 2 normally done when delivered RS 232 PIN NUMBER PIN NAME 2 RXD 3 TXD 5 GND 7 TARGET TTL 5V OV 8 TTL 5V Set the target weight...

Отзывы: