Introduction

Transcell new Weighblok series are a rugged digital scale casted in strong aluminum. It comes standard with a large (.75") LCD screen for easyreadout.
All setup parameters may be entered via the membrane panel keys. The Scale has an

intelligent

auto power off function and can be operated

with 6 alkaline 1,5 V batteries. If the battery is not good enough to ensure a correct value the scale will turn off.

The scale is developed especially for weighing of refrigerant cylinders, and to control when transferring refrigerants to other cylinders or equipments.
The scales setup parameters are altered through the Setup menu while a few other parameters are altered through the User menu The
configuration section of the manual explains how to use the five front panel keys to maneuver and save settings in both menus.

The Weighblok scale can be used as a Batching Scale or regular weighing scale

The Weighblok is suitable for use in weighing applications as: refrigeration cylinders, fire protection cylinders, CO2,
cylinders, parcels, persons, etc.

Keyboard functions
The calibration switch is only for calibration.

Indicator ON
In not EC-type approved version: Sends "Print" data to printer if scale is stable and not in overload.
Not active when "Continuous" option is selected in User Menu.

Dot

. Extended graduation when pushed and 5 sec after release. When pressed and released, the DOT

button is the unit switch key, lb

kg.

Toggles between Gross and Net weight display only if a Tare has been established.

Sets scale to display "0" when in Gross mode, and within zero band range. When pushed in for 3
seconds, the Scale switches off.

Used to establish a Tare (zero the Scale) while in either Gross mode or
Net mode. This operation cannot be performed at or below Gross zero.

ON

PRINT

NET

GROSS

OFF

ZERO

TARE

Содержание WEIGHBLOK

Страница 1: ...Revision 0502 4 2 03 2001 2002 Transcell Technology Inc Contents subject to change without notice Transcell Technology Inc 975 Deerfield Parkway Buffalo Grove IL 60089 Tel 847 419 9180 Fax 847 419 15...

Страница 2: ...Manual Normal Calibration Calibration switch ON PRINT NET GROSS OFF ZERO TARE Communica tions port External power 9 VDC Alkaline batteries The Weighingblock is a robust Scale It is designed for portab...

Страница 3: ...enu while a few other parameters are altered through the User menu The configuration section of the manual explains how to use the five front panel keys to maneuver and save settings in both menus The...

Страница 4: ...ve to the selection value press ZERO once 7 Increase and decrease the value with PRINT and TARE 8 To save the value press NET 9 Press DOT key to go back to parameter number 10 To go back to normal men...

Страница 5: ...accuracy The higher the filter number the greater the accuracy but the slower the response time Choose 4 or 8 unless a very fast response is needed 1 2 4 8 F7 Overload Limit Selects the desired formul...

Страница 6: ...field Parkway 800 503 9180 www transcell net Buffalo Grove IL 60089 847 419 1515 transcell transcell net F6 F7 F8 F9 A10 Auto Power Off Period Selects the auto power off time period in minutes Off Dis...

Страница 7: ...omplete indicator calibration To adjust the max point using the F17 span adjusts procedure 1 Scroll to F 17 then press ZERO to enter span calibration menu 2 The display will momentarily show C 1 for t...

Страница 8: ...erating Mode Indicates a low battery condition Re charge the battery or replace if alkaline batteries Err 0 Span Calibration Mode F17 Keyed in weight value is larger than full scale capacity Use a sma...

Страница 9: ...normal mode Loadcell Connections Check the Loadcell with a universal voltmeter see table The left pin is number 1 Terminal J1 Description Colours Loadcell AG Colours Loadcell 1042 E 1 Out 5 VDC Brown...

Страница 10: ...appears Basic scale use 1 Press ON to start the scale 2 If necessary press ZERO to obtain a weight reading of zero 3 Press OFF to switch off the scale Hold 3 seconds 4 The scale switches off automatic...

Страница 11: ...when filling or discharging One time set up F11 must be 1 and A5 must be 2 normally done when delivered RS 232 PIN NUMBER PIN NAME 2 RXD 3 TXD 5 GND 7 TARGET TTL 5V OV 8 TTL 5V Set the target weight...

Отзывы: