Calibration
26
RMWH-SVX001C-EN
Stopping Calibration
During some phases, when
Cancel
is available in the top
left corner, the calibration process can be stopped.
1. To stop the calibration process, select
Cancel
.
2. For zero-gas calibrations, disconnect the zero-gas
scrubber tubing from the calibration port, then select
Next
on the device touchscreen.
3. On the
Canceled
screen, select
Done
.
4. On the bottom of the device, install the pin in the
calibration port.
Zero Gas Calibration
1. Connect the zero-gas scrubber tubing to the calibration
port.
2. On the device touchscreen, select
Next
.
A progress bar shows during the zero gas calibration.
3. When zero gas calibration is complete, the
Results
screen shows. Select
Continue
to go to the next step.