User Manual
TRAQC-8 DPC
TRAQC-8 DPC 11/2016 EN Rev.3
Sheet 28
4.3
Electrical Output A Setup
(optional)
In the main menu select “Electrical Output Setup”.
The following options will be displayed:
Select with ‘ARROW’ keys as required, press ‘ENTER’ key (F1):
4.3.1
Pressure xmtr mode (Pressure transmitter mode)
The TRAQC-8 DPC will be used to replace a pressure transmitter output for calibration purposes.
Select:
mA xmtr mode.
mA Source
On/Off
mA simulator
V xmtr mode
On/Off
In “mA source” the TRAQC-8 DPC will supply electrical power and control the mA output. (effectively a
controlled current source).
In mA simulator an external electrical power supply is required, and the TRAQC-8 DPC will only control the
mA output.