User Manual
TRAQC-8 DPC
TRAQC-8 DPC 11/2016 EN Rev.3
Sheet 30
4.3.5
Voltage Output
The Voltage Output option is identical to the “pressure Xmtr” mode, with exception of the output setting (as
set in the next step:” Voltage output Settings”) : Output setting will not be in pressure units, but in Volt.,
with fixed Volt steps to go from minimum to maximum. Voltage output will switch on / off in this menu.
4.3.6
Voltage Output Settings
As explained in “Xmtr mode settings” there are no units to select: The
user is requested to enter Lowest value, highest value and step value
direct in V. Values in excess of the standard range of 0-11 V will be
refused: when ENTERED the instrument will refuse and display: “
Invalid entry”. The display will wait for a valid entry before proceeding
to next entry
Following functions are only available when “Current output” or “Voltage output”
has been switched to “on”.
4.3.7
Auto Span Check Mode
Select “On”, automatically the “step time entry” panel is displayed: Enter
the required step time using the ‘ARROW’ keys, then press ‘ENTER’ key
(F1).
Select “off ”: the function has been disabled.
4.3.8
Ramp Mode
Select “On”, automatically the “Number of steps entry” panel is displayed:
Enter the required number of steps using the ‘ARROW’ keys, and then
press ‘ENTER’ key (F1).
Automatically the “Step time entry” panel is displayed: Enter the required
step time in seconds, using the ‘ARROW’ keys.
Press ‘ENTER’ key (F1) to confirm.
Select “off”: the function has been disabled.
4.4
Electrical input B setup (optional)
This option is under development
4.5
Electrical output B setup (optional)
This option is under development