User Manual
TRAQC-8 DPC
TRAQC-8 DPC 11/2016 EN Rev.3
Sheet 22
4.1.3
Averaging on/off
A progressing averaging can be applied. The reading speed is standard three times per second. The objective
is to display the fluctuating measuring signals in a reliable manner. If averaging is switched on, this is shown
by ‘
𝑥̅
’ above the pressure unit reading in the ‘measuring mode’ menu.
To enable or disable the averaging function select the “averaging on/off” option in the “pressure setup” menu.
The default state is the “averaging off”. To change the current state press ‘ENTER’ key (F1).
4.1.4
Averaging setup
The number of times that averaging must be performed can be set as
follows: Select the “averaging setup” in the pressure setup menu with the
‘ARROW’ keys and press ‘ENTER’ key (F1) to confirm. Now use the ‘ARROW’
keys to change the number of cycles to the desired one, press ‘ENTER’ key
(F1) to confirm.
The number of cycles default is 6. The cycles can be set from 3 to 9.
4.1.5
Resolution Setup
To change the number of digits (from 3 to 6) select “resolution setup” in the
“pressure setup” menu and use the ‘ARROW’ keys to change the number of
digits, confirm by pressing the ‘ENTER’ key (F1)