MD1064 Rev. F
27 of 32
4
channels per Optimus™ control module. Therefore, up to 72 stages are available
depending on system configuration.
Bit
Description
0
Stage 1 Failed
1
Stage 2 Failed
⁞
⁞
n
Stage (n + 1) Failed
⁞
⁞
71
Stage 72 Failed
Table 16. Failed stages bit field description (Bytes 0-8).
Byte
Format
Description
0-8
Bits (72)
Stage Failures
Table 17. Command 160 response data bytes.
Code
Class
Description
0
Success
No Command Specific Errors
Table 18. Command 160 specific response codes.
6.2.4.4
Command 166
–
Read Radiography Status
This device specific command provides a mapping of failed stages that are reporting
radiography within the system. The following tables describe the construction of the
command.
Bit
Description
0
Stage 1 Detecting Radiography
1
Stage 2 Detecting Radiography
⁞
⁞
n
Stage (n + 1) Detecting Radiography
⁞
⁞
71
Stage 72 Detecting Radiography
Table 19. Radiography detect bit field description (Bytes 1-9).
Byte
Format
Description
0
U8
Number of Detected Radiography Events
1-9
Bits (72)
Radiography Detect Stages
Table 20. Command 166 response data bytes.
Code
Class
Description
0
Success
No Command Specific Errors
Table 21. Command 166 specific response codes.