MD1064 Rev. F
17 of 32
The HART “squawk” feature has been implemented in the Optimus™ control module. A
single squawk sequence will be the same as the test sequence performed on power up.
Continuous squawking will be the same sequence repeated.
PRI150 Discovery Mode
To indicate that the T221 control module is in PRI150 Discovery Mode, LED 1 (power and
status) will alternate between RED and YELLOW at 1Hz (i.e. 500ms RED, 500ms YELLOW).
LED
State
Description
PWR
Red/Yellow In PRI150 Discovery Mode
CH1-4
1Hz Green
Channel Currently Being Discovered
Green
Channel Discovered
Yellow
Channel Yet to be Discovered
Off
Channel Not to be Discovered
HART
Green
T221 Communicating with Host
Yellow
Host Communicating with T221
Off
No Ongoing Communications with Host
Table 4. Local display during PRI150 Discovery Mode.
After the last PRI150 channel has been discovered, the device will stay in PRI150 discovery
mode for a further 1 second. This is to allow the end user to see that the channel has
successfully been discovered.
4.3
Software Installation
To fully utilise the Optimus™ calibration and advanced diagnosti
cs, the Tracerco ToolBox
software must be installed. Refer to the Tracerco ToolBox documentation.
Currently, there is no HART device descriptor available. However, the necessary HART
details are provided to implement advance functionality in your DCS. Refer to Tracerco
Document TD1071 T221 HART Field Device Specification for complete details.
Содержание Optimus T221
Страница 1: ...Author Paul Storey Date 10 December 2019 Document no MD1064 F Operation Manual T221 Optimus ...
Страница 32: ......