Configuration / Parameter setting via the Interbus-S master
©
TR-Electronic GmbH 2007, All Rights Reserved
Printed in the Federal Republic of Germany
Page 26 of 37
TR - ELE - BA - GB - 0008 - 06
14.12.2007
6 Configuration / Parameter setting via the Interbus-S master
The configuration of the laser occurs alternatively via the configuration software of the
Interbus-S - master or via the TRWinProg-software. With a download of the control
parameters the parameters, which were configured via the TRWinProg-software, will
be overwritten by the control.
In this instruction only the configuration via the Interbus-S - master is described. The
PC program TRWinProg is described in an instruction of its own.
6.1 Parameter data
The parameter data can be transferred via the bits 0 to 24 of the process out data
channel of the master to the laser. To this the laser must be set to the
parameterization state. This is achieved by outputting a parameter number unequal to
zero on bits 9 to 12 of the control word (bits 25 to 28 of the process out data channel).
Parameter overview
Parameter No.
B12 B9
Function
0 0 0 0
Output in "operation" state
0 0 0 1
Resolution:
0 = 10 mm
1 = 1 mm
2 = 0,1 mm
3 = 0,01 mm
4 = 1 Inch
5 = 0,1 Inch
0 0 1 0
Switch off laser diode
0 0 1 1
Switch on laser diode
0 1 0 0
Preselection preset value
0 1 0 1
Automatic switching-off of the laser diode, if bus 15min. inactively:
0 = disabled
1 = active
0 1 1 0 Not
defined !
0 1 1 1 Not
defined !
1 0 0 0
Counting direction:
0 = with increasing distance to the laser, values increasing
1 = with increasing distance to the laser, values decreasing
1 0 0 1
Speed limit value in 0,1m/s:
0 = no check
1 – 200 = 0,1 m/s – 20 m/s
1 0 1 0
Error value at beam interruption:
0 = NULL
1 = 0xFF
2 = Last valid position value
1 0 1 1 Not
defined !