Page 5
Press to record readings into memory. See section 11.1. Also used
to switch the Altitude or Atmospheric Pressure Correction system on
or off. See section 9.1.
Press to alternate temperature, date/time, and manual
Salinity/Altitude/ Pressure values on the display. See section 16.2.
Also used to select pH6.88 or pH7.00 as the primary pH buffer
(section 17).
Press to start or stop automatic logging. See section 12.
Alternatively, press to transmit current reading plus date and time to
the RS232 port (optional) See section 13.2.
Only used within the menu system on the
WP-91
.
Press to access the user-friendly menu system which makes the
WP-91
easy to operate.
and
The
and
keys are used for calibrating temperature readout
(section 7.1), setting the clock (section 16.1), setting the automatic
logging period (section 12), and displaying GLP information (section
10.1). The
key is also used to initialise the
WP-91
at turn-on. See
section 18.
Switches the
WP-91
on and off.
Display
32 character alpha-numeric display with user-friendly menu and
prompting system. Shows Dissolved Oxygen, pH/mV and
Temperature simultaneously. Date, time, manual salinity value,
altitude or pressure can also be displayed.