Page 40
13.7 GLP Data Format
GLP information is returned as up to 8 lines terminated by a carriage
return. When using the “
?G
” command (section 13.5), the computer must
respond with a character after receiving each line. For example…
WP91 V1.2 S1234 @ 31/12/03 12:00
Oxygen Zero= 0.0% @ 31/12/03 11:00
Oxygen Span= 100.0% @ 31/12/03 11:10
Oxygen Altitude= 5000m @ 31/12/03 11:10
Oxygen Pressure= 1013HPa @ 31/12/03 11:10
pH Asy= 0.10pH @ 31/12/03 11:20
pH Slope= 100.0% @ 31/12/03 11:30
Temperature Offset= 1.0oC @ 31/12/03 11:40
ENDS
Notes:
1. Either Altitude
OR
Pressure is sent, depending upon which was
selected the last time the instrument was calibrated. If the Altitude or
Pressure Correction was switched off at the time of the last air
calibration, then neither is displayed.
2. pH Asymmetry and pH Slope calibration data are only sent when the
pH mode of the
WP-91
is switched on.