Page 19
5.2 Calibration Notes
1. If a 1-point calibration has been successfully performed, the
WP-91
will display the following message, and the asymmetry of the sensor.
1 Point Cal. OK
Asy= 0.10pH
2. If a 1-point calibration has failed, the
WP-91
will display the following
message, and the failed asymmetry value of the sensor.
1 Point Cal.Fail
Asy= 1.50
or
1 Point Cal.Fail
Asy=-1.50
3. If a 2-point calibration has been successfully performed, the
WP-91
will display the following message, and the asymmetry and slope of
the sensor.
2 Point Cal. OK
Asy= 0.10pH
then
2 Point Cal.OK
Slope=100.0%
4. If a 2-point calibration has failed, the
WP-91
will display the following
message, and the failed slope value of the sensor.
2 Point Cal.Fail
Slope=130.0%
or
2 Point Cal.Fail
Slope= 70.0%
6. mV Calibration
The mV section is factory calibrated. There is no user-calibration facility
for this mode.