Page 31
11. Notepad Function
11.1 Recording Readings into the Notepad
To record readings into the Notepad memory:
1. Press
in normal display mode. The display should now look like
this…
10.00
M 7.00
F1
:
1 12:00:00
2. If you now press
, the Dissolved Oxygen, pH/mV, Temperature,
Date and Time will be recorded into the notepad, and labelled as
reading number 1.
If manual salinity, altitude or pressure correction are in use, they are
also recorded with the reading.
3. Repeat steps 1 & 2 as often as required. The maximum number of
readings that can be stored in the Notepad is 150.
11.2 Recalling Records from the Notepad
To recall records from the Notepad onto the
WP-91
display:
1. Select the Notepad menu (
→
F2:Notepad
).
2. Select
F1: Recall
from the menu.
3. Record number 1 is now displayed.
An “
s
” is displayed if manual Salinity correction was switched on
when the reading was recorded.
An “
A
” is displayed if Altitude correction was switched on when the
reading was recorded.
A “
P
” is displayed if Pressure correction was switched on when the
reading was recorded.
For example…
10.00
MsA 7.00
25.0
o
c 1 F2
:
Clk
or
10.00
MsP 7.00
25.0
o
c 1 F2
:
Clk