Page 22
8. Salinity Correction
Manual salinity correction for ppM Dissolved Oxygen readings is
available on the
WP-91
.
1. Select Salinity-corrected ppM mode.
(
→
F3:Mode
→
F1:ppM
→
F2:ON
).
2. Set the Salinity correction value. (
→
F1:Cal
→
F4:Sal
).
If Altitude of Pressure Compensation is currently selected, then the
above key sequence becomes
→
F1:Cal
→
F4:Comp.
→
F1:Sal
3. The current salinity correction value is now displayed…
> 36.0< ppK Sal.
↑↑↓↓
:Set F1:Save
4. Press the
and
keys until the display shows the desired salinity
correction value.
5. Press the
key to save the salinity correction value.
Alternatively, press the
key to quit and retain the current setting.
6. When manual Salinity correction is selected, an “
s
” is shown after the
“
ppM
” units description in the Dissolved Oxygen readout.