
Page 45
20.3 pH Temperature Compensation
The slope of a pH sensor (section 20.2) is affected by temperature. This effect is
compensated for either by using an Automatic Temperature Compensation (ATC)
probe or by entering the sample temperature manually. Figure 20-3 shows the
slope of a pH sensor at various temperatures.
-600
-400
-200
0
200
400
600
0
7
14
Millivolts
Electrode
Potential (mV) at
0 oC (54mV/pH)
Electrode
Potential (mV) at
50 oC (64mV/pH)
Electrode
Potential (mV) at
100 oC
(74mV/pH)
pH Electrode Response, as a Function of Temperature
pH
Figure 20-3
20.4 Checking the reference junction of a pH sensor
If pH readings are inaccurate or unstable, the reference junction of the sensor may
be blocked. The following test can be performed to determine if the reference
junction of a pH sensor is making adequate contact with the sample solution.
1.
Calibrate the
WP-90
, as per section 5.1.
2.
Dilute 1 part of pH6.88 buffer with 9 parts of distilled water.
3.
Measure the pH of the diluted buffer. The result should be 7.06 +/-0.02 pH.
4.
If the value obtained is outside of these limits, then clean the reference
junction, as per the instructions supplied with the pH sensor.
5.
Re-calibrate the
WP-90
and repeat the test.
6.
If the value obtained is still outside 7.06 +/-0.02 pH, then the sensor should
be replaced.