Page 32
12.5 Commands
The following commands can be sent from a PC to the
WP-90
. Note that <cr>
denotes carriage return and <lf> denotes a line feed.
Action
Command
Notes
Request current data
?D<cr>
Returns the current Specific Ion/pH/mV,
Temperature, date and time from the
WP-90
. The log number returned is set
to zero.
Request logged data
?R<cr>
Returns all logged records from the
WP-90
memory. The data ends with the
message
ENDS
<cr>
Erase logged data
?E<cr>
Erases all logged records from the
WP-
90
memory. Returns the message
ERASED
<cr> to confirm that the
records have been erased.
Request status
information
?S<cr>
Returns the model name, firmware
version number, instrument serial
number and number of logged readings
in memory,
eg:
WP90
ss
V1.0
s
R1234
s
9999
<cr>,
where
s
are spaces. Note that the
number of logged readings is right-
justified.
Request GLP
information
?G<cr>
Returns all calibration GLP information,
plus the instrument model and current
date (see section 12.6 for data format
and handshaking).