Page 33
12.6 Data Format
Data is returned to the RS232 port by the
WP-90
in the following format…
LLLL
ss
DDDDDDDDUUU
ss
TTTTTTTTuuu
ss
dd/mm/yy
ss
hh:mm:ss
where…
LLLL
is the Log Number. Maximum 4 characters, right justified.
The
WP-90
sends a Zero for instant readings (section 12.2)
ss
is one space.
DDDDDDDD
is Specific Ion, pH, mV or Relative mV data, maximum 8
characters, right justified.
UUU
is the unit description. Either “
ppM
”, “
ppK
”, “
pH
ss
”
, “
mV
ss
”,
“
mVR
”. “
sss
sss
” is sent for Exponential readout of Specific Ions.
ss
is one space. (
ss
are spaces.)
TTTTTTTT
is the Temperature Data, maximum 8 characters, right justified.
uuu
is the Temperature unit description. Either “
oC
ss
” for real
temperature data (where
ss
is 1 space), or “
oCm
” for manual
temperature compensation data.
ss
is one space.
dd/mm/yy
is the date, month and year data.
ss
is one space.
hh:mm:ss
is the hours, minutes and seconds data.
When data is requested by a PC with the ?D or ?R commands (section 12.5), it is
terminated with a carriage return.
When data is sent by the
WP-90
using the Print function (section 10.4) or the
Instant Send function (section 12.2), the data ends with a carriage return and a line
feed.