17. Initialising the WP-90
If the calibration settings of the
WP-90
exceed the allowable limits, the unit may
need to be initialised to factory default values.
To initialise the
WP-90
…
8.
Switch the
WP-90
OFF
.
9.
Press and HOLD the
key while switching the
WP-90
back on.
The following messages should be displayed…
Initialized
MUST ReCalibrate
then:
WP90s V1.0 R1234
Ion/pH/mV & Temp
(The “s” after
WP-90
is shown when the RS232 serial port option is fitted)
10.
The meter then displays Specific Ion and Temperature. Note that the decimal
points have been replaced with a
∗
, to indicate that the unit requires re-
calibration.
Note:
When the
WP-90
is initialised, the primary pH buffer is re-set to pH6.88 and
the high pH buffer is re-set to pH9.23. See section 16 if you wish to use
pH7.00 and/or pH10.00 buffers.
18. Instrument firmware version number.
If you need to phone or fax TPS for any further technical assistance, the version
number of your
WP-90
firmware may of benefit to us. The version number is
displayed by the
WP-90
at turn-on. In the example below, the firmware version
number is
V1.0
.
WP90s V1.0 R1234
Ion/pH/mV & Temp