Page 12
4.
Enter the value of the Primary Standard to be used for calibration.
1.000ppM Pri Std
F1
:
←
←
F2
:
→
→
↑↑↓↓
:Set
Use the
and
keys to set the first digit. The numbers scroll from 0
to 9 plus the decimal point.
Press
to move to the next digit, then use the
and
keys to set
it.
Continue moving from digit to digit until the value of the Primary Standard
has been correctly set. Up to 4 digits plus a decimal point can be entered.
After the last digit has been entered,
will move the cursor to the units.
Use the
and
keys to select ppM or ppK.
Press
to save the Primary Standard.
5.
Enter the value of the Secondary Standard to be used for calibration.
The Secondary Standard must be at least 2 to 100 times higher or lower than
the Primary Standard.
10.00ppM Sec Std
F1
:
←
←
F2
:
→
→
↑↑↓↓
:Set
Use the
and
keys to set the first digit. The numbers scroll from 0
to 9 plus the decimal point.
Press
to move to the next digit, then use the
and
keys to set
it.
Continue moving from digit to digit until the value of the Secondary Standard
has been correctly set.
Up to 4 digits plus a decimal point can be entered.
After the last digit has been entered,
will move the cursor to the units.
Use the
and
keys to select ppM or ppK.
Press
to save the Secondary Standard.
6.
The
WP-90
will now return to normal measurement mode.
7.
A “ * ” will be shown is place of the decimal point if the Specific Ion mode is
not calibrated, or if calibration has failed.