Page 9
Memory .................................... 3600 readings including date and time
Automatic Logging....................
User-set for one reading every 1 to 90
seconds, 1 to 90 minutes, or 1 to 24 hours.
RS232 Output ......................... 300, 1200, 9600 & 19200 baud. 8 bits, no
parity, 1 stop bit, XON/XOFF Protocol.
Clock ........................................ Calendar clock displays date, month,
hours, minutes & seconds.
Year is recorded in memory and
transmitted to the RS232 port, but is not
displayed.
Battery Saver ........................... On : Auto switch-off after 5 minutes
Off : Continuous use
Bar Graph display of battery charge level.
Readout of battery voltage available for
troubleshooting.
Good Laboratory Practices ....... Date, Time and Value of last pH and
Temperature calibration are stored, and
can be recalled or sent to the RS232 port
at any time.
Power ....................................... 6V NiMH Rechargeable Battery for
approx 50 hours operation.
Dimensions .............................. 195 x 110 x 55 mm
Mass ........................................ Instrument only : Approx 520g
Full Kit
: Approx 2.5kg
Environment ............................. Temperature
: 0 to 45
o
C
Humidity
: 0 to 95 % R.H.