Page 17
8. When the reading has stabilised, press the
key to zero the Relative
mV/ORP reading. The Relative mV/ORP reading will now be zero, and
the absolute mV/ORP reading will remain unchanged. For example…
0.0R 7.00pH
660
F4:Zeros
or:
0.0R 7.00pH
31/12 12:00:00
9. The
WP-80D
Relative mV/ORP mode is now zeroed and is ready for
use. The readout can be re-zeroed by pressing the
key whenever
required.
6.1 Notes
1. If Relative mV/ORP is selected for
BOTH
channels, then the
WP-80D
offers you a choice between Channel 1 or Channel 2 when you press
the
key to zero the Relative mV/ORP reading.
Select Channel
F1:Chan1 F2:Chan2
2. The Relative mV/ORP offset is retained in memory when the
WP-80D
is switched off, even when the battery is removed.
3. The Relative mV/ORP zero is reset when leaving Relative mV/ORP
mode (see section 3)